StagingMerge - DEN-17260-ui-dn-release-21 (13 commits squashed) DEN-17260: UI-DN-Release-21 - DEN-17261: Corrected the treatment logs doxygen and data type for potassium and calcium.
DEN-17260: UI-DN-Release-21
- DENBUG-298: Removed the Maximum Saline Bolus State - DENBUG-298: Added Deliver Saline Bolus in treatment stop.
DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating
- Reduced the font size of fluid value font from 32 to 28.
DEN-17269: Change_version_format_from_dot_to_dash
- Changed the format of the version from dot to dash before compatibility value.
DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same
- Modified message 0xBA to send up to 4 float payloads to the UI - Changed MDuetConfirmHD to receive and process the new payloads - Changed VConfirm to update the confirm messages that have {} in them and replace them with actual values
DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating
- Updated the TreatmentSaline.qml file to clear the reject reasons associated with saline bolus once the treatment state has changed.
DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same
- Updated the VConfirm to trigger the confirm messages even if the ID has not changed.
DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating
- Updated the code to clear the saline bolus related reject reasons upon changing the saline bolus states.
DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating
- Added code to clear the saline bolus reject reason upon changing the treatment states.
DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same
- Updated the code to round up the UF rate that is displayed on the UF rate change screen.
DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same
- Updated the current UF rate in the VTreatmentUltrafiltration to be rounded up. - Removed commented code.
DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same
- Made UF Rate related values in VTreatmentAdjustmentUltrafiltrationEdit to be rounded up prior to displaying on the UI.
DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same