Clone
Dara Navaei
committed
on 18 Mar
StagingMerge - DEN-17260-ui-dn-release-21 (13 commits squashed)
DEN-17260: UI-DN-Release-21
- DEN-17261: Corrected the treatment logs doxyg… Show more
StagingMerge - DEN-17260-ui-dn-release-21 (13 commits squashed)

DEN-17260: UI-DN-Release-21

- DEN-17261: Corrected the treatment logs doxygen and data type for potassium and calcium.

DEN-17260: UI-DN-Release-21

- DENBUG-298: Removed the Maximum Saline Bolus State

- DENBUG-298: Added Deliver Saline Bolus in treatment stop.

DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating

- Reduced the font size of fluid value font from 32 to 28.

DEN-17269: Change_version_format_from_dot_to_dash

- Changed the format of the version from dot to dash before compatibility value.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Modified message 0xBA to send up to 4 float payloads to the UI

- Changed MDuetConfirmHD to receive and process the new payloads

- Changed VConfirm to update the confirm messages that have {} in them and replace them with  actual values

DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating

- Updated the TreatmentSaline.qml file to clear the reject reasons associated with saline bolus once the treatment state has changed.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Updated the VConfirm to trigger the confirm messages even if the ID has not changed.

DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating

- Updated the code to clear the saline bolus related reject reasons upon changing the saline bolus states.

DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating

- Added code to clear the saline bolus reject reason upon changing the treatment states.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Updated the code to round up the UF rate that is displayed on the UF rate change screen.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Updated the current UF rate in the VTreatmentUltrafiltration to be rounded up.

- Removed commented code.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

-  Made UF Rate related values in VTreatmentAdjustmentUltrafiltrationEdit to be rounded up prior to displaying on the UI.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Addressed the code review comments

Show less

release/p1b/master + 5 more