DEN-15664: Blood Pump Stop Time Exceeded Alarms (124, 125, 165, and 212) Alarm Timer Feature is Inconsistent - Added the txEnd to the condition as well to display the countdown timer.
DEN-13474: Updated - UI: SRSUI for SD-Card/File System Write Failure & Export Log Failure may be Outdated from PRS (SRSUI 1077-1079)Rename the signal/slot per CR request
DEN-15805: Updated - UI: Pressure Adjustment Screen Does Not Reset When User Cancels ChangeModify fix to use similar technique for setting and resetting in treatment flow screen
DEN-12842: Updated - UI: prepopulate option on the create treatment screen (Phase 1b)Remove model refactoring and restored to originalCorrected slider + switch state update for prepopulating
DEN-15769: UI: Usability: Remove Option to increase/decrease treatment time after UF volume adjustments - Fixed the added new state from common develop (blood recover)