Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

    • -237
    • +219
    /firmware/App/Modes/ModeFill.c
DEN-15367 changed the RO permeate sample cancel water path. DEN-14004 changed the request message from UI to 0x89. DEN-15509 changed the reservoir threshold from 1700 to 1850 mL. DEN-15192 changed the concentrate pump low speed threshold.

    • -5
    • +10
    /firmware/App/Modes/ModeROPermeateSample.c
Merge branch 'DEN-15367-hd_dg_test_configuration_setup' of ssh://devapps.diality.us:7999/dg/dgfirmware into DEN-15367-hd_dg_test_configuration_setup

-DEN-15367 fixed the transition to cancel water path in RO permeate sample mode -DEN-14004 changed the request message ID to 0x89 from UI to send the DG service record -DEN-15192 changed the concentrate pumps low speed check threshold -DEN-15509 changed the reservoir full threshold from 1700 mL to 1850 mL.

DEN-14004 added another message to receive and send DG service record upon request from HD. The format of the payload is different from the format that is sent to UI

DEN-15367, DEN-15389, DEN-15511, DEN-15443 added code in the mode fill to check for the conductivity of the values in the produce state and have a timeout. Removed the acid check at the end of the mode fill

    • -237
    • +219
    /firmware/App/Modes/ModeFill.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-15367

DEN-15367, DEN-15404 worked on RO permeate sample mode

    • -7
    • +12
    /firmware/App/Modes/ModeROPermeateSample.c
DEN-15367, DEN-15404 work in progress. Updated the RO permeate sample mode code to be able to receive and send ready command to HD for further dispensing or canceling the run

    • -54
    • +44
    /firmware/App/Modes/ModeROPermeateSample.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

DEN-15367 added test configuration to be able to run chemical disinfect without PAA acid

DEN-15367, DEN-15446 renamed the function that checks whether any of the inlet water conditions are active in a cleaning mode

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-15367

    • -2
    • +14
    /firmware/App/Services/SystemComm.c
DEN-15446 added a function to check the status of the inlet water conditions and trigger the cleaning modes to transition to cancel water path

    • -3
    • +6
    /firmware/App/Modes/ModeHeatDisinfect.c
Merge branch 'staging' into DEN-15367-hd_dg_test_configuration_setup

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-15367 merged staging

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-15478

DEN-15478 updated the code to not check for the integrated volume vs. flow sensors if the fill is less than 550 mL

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo