hdfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 changed the location of the saline bolus request in treatment stop

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 added code in saline bolus to hold the start flag to true in treatment stop so the treatment stop does not transition

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 added code to accept saline bolus request in tretment stop as well

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 added the treatment mode to the minimize code in alarm management

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DENBUG-298

DENBUG-299 added more comments around the disabled code for release 21

DENBUG-298: Addressed the code review comment. DENBUG-299: Disabled the code for release 21

DENBUG-298 removed the maximum volume in the rinseback additional

DENBUG-298 moved the saline bolus request to the top of the treatment stop alarms and signals function

DENBUG-298 addressed the code review comments. Removed the maximum saline bolus check code

    • -53
    • +10
    /firmware/App/Modes/SalineBolus.c
DENBUG-298 initial commit of the SalineBolus module. Changed the code based on the new saline bolus code

    • -389
    • +41
    /firmware/App/Modes/Dialysis.c
    • -0
    • +445
    /firmware/App/Modes/SalineBolus.c
    • -0
    • +53
    /firmware/App/Modes/SalineBolus.h
    • -2
    • +33
    /firmware/App/Modes/TreatmentStop.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-311 changed the speed of persistence rotor speed for alarm to 20 seconds

    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
DENBUG-302, DEN-17243 added code to set the UF diff volume to 0 if the negative in treatment duration change confirmation

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-311 changed the persistent time of alarm 38 to 20 seconds

    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DENBUG-298