DENBUG-302 added more comments

DENBUG-302 changed the UF rate that is reported to UI in mL/min

DENBUG-302 updated the code to check for the UF rate confirmation in operation mode. DEN-17237 changed the code to re-trigger the placeholder cleaning mode alarms if they got cleared

  1. … 1 more file in changeset.
Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

  1. … 3 more files in changeset.
DENBUG-302 continued with the changes for the UF rate screen support

  1. … 3 more files in changeset.
DENBUG-302 added code to ignore sending the UF rate to the user if the UF volume is 0

DENBUG-298 addressed the code review comment

DENBUG-302 updated the code to send the UF rate back to the user and wait for the user to accept or reject it

  1. … 1 more file in changeset.
DENBUG-298 removed code to not countdown when in saline bolus and in treatment stop

DENBUG-302 changed the code to send the new calculated UF rate to the user to accpet or reject and then based on that the treatment duration is updated

  1. … 2 more files in changeset.
DENBUG-298 changed the block saline bolus to rinse back from dialysate recirculation

  1. … 2 more files in changeset.
DENBUG-298 addressed the code review comment

  1. … 4 more files in changeset.
DENBUG-298 addressed the saline bolus rejection conditions

DENBUG-298 added code in saline bolus to reject saline bolus request if the active alarm has blood and dialysate recirculation blocked

DENBUG-298 fixed the 4th level for saline bolus in treatment stop

DENBUG-298 updated the code

DENBUG-298 addressed the code review comments

DENBUG-298 added broadcast to saline bolus in treatment stop. Made the stop bolus available in treatment stop too. Made VBV to open in saline bolus

DENBUG-298 changed the location of the saline bolus request in treatment stop

DENBUG-298 added code in saline bolus to hold the start flag to true in treatment stop so the treatment stop does not transition

DENBUG-298 added code to accept saline bolus request in tretment stop as well

DENBUG-299 added more comments around the disabled code for release 21

DENBUG-298: Addressed the code review comment. DENBUG-299: Disabled the code for release 21

DENBUG-298 removed the maximum volume in the rinseback additional

DENBUG-298 moved the saline bolus request to the top of the treatment stop alarms and signals function

DENBUG-298 addressed the code review comments. Removed the maximum saline bolus check code

DENBUG-298 initial commit of the SalineBolus module. Changed the code based on the new saline bolus code

  1. … 2 more files in changeset.
DENBUG-302, DEN-17243 added code to set the UF diff volume to 0 if the negative in treatment duration change confirmation

DENBUG-298 addressed the code review comments

DENBUG-302 updated the code to change the UF rate instead of the UF volume upon treatment duration change request