DEN-15893-cartridge-lever-is-open-alarm-when-cartridge-lever-is-closed

Cleared alarm 120, prior to initiating alarm 85

Required latch closed before allowing alarm 85

Paused in WET self test until bubbles clear or latch closed.

DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-2

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge remote-tracking branch 'origin/DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus' into staging

DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to

Suspenders and belt. no rinseback/ no recirculate required to be true.

DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to

Alarm only suppressed when end treatment is the *only* option, no OK and continue

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15664-blood-pump-stop-time-exceeded-alarms-124-125-165-and-212-alarm-timer-feature-is' into staging

Fixed. Thank you!

Fixed. Thank you!

DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to

Fixed typo.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15664-blood-pump-stop-time-exceeded-alarms-124-125-165-and-212-alarm-timer-feature-is' into develop

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

The trimmer heater is turned back on in the transition back to dialysis: transitionToDialysis( ) This is definitely a place I would suggest refactoring, as the trimmer heater should follow dialysa...

The trimmer heater is turned back on in the transition back to dialysis: transitionToDialysis( )

This is definitely a place I would suggest refactoring, as the trimmer heater should follow dialysate flow everywhere, but is somewhat distributed.

Where is the heater turned back on? Is the state saved?

Where is the heater turned back on? Is the state saved?

Is this intended to be a BITWISE OR? Did you intend || ?

Is this intended to be a BITWISE OR? Did you intend || ?

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to' into staging

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge remote-tracking branch 'origin/DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus' into develop

DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus

Stop the pumps more forcefully, open loop, prior to switch valves for the saline bolus.