Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DEN-15628

DEN-15628, DEN-15515 fixed the zero sequence hang up issue. Increased the number failed blood leak command from 3 to 10 commands

    • -11
    • +11
    /firmware/App/Controllers/BloodLeak.c
DEN-15628 addressed the blood leak zero sequence intermittent hang up

DEN-15788: Increased open loop DPo start rate for dialysate/dialyzer prime state.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15788-pbo-ramps-up-to-400-mmhg-during-dialysate-circuit-priming' into develop

DEN-15788: open VBA and VBV when checking for atmosphere pressure before pressure leak test.

Merge branch 'staging' into DEN-15716-skipped-treatment-start-state-while-transitioning-from-pre-treatment-to-treatment

    • -24
    • +124
    /firmware/App/Modes/SelfTests.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15788-pbo-ramps-up-to-400-mmhg-during-dialysate-circuit-priming' into develop

DEN-15788: change DPo control during dialysate/dialyzer prime state to begin immediately instead of after fixed prime time.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused

force pumps to stop in recirc/treatment mode, on a manual stop.

DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to

Retrigger blocked, alarms directly blocked during stop

    • -7
    • +11
    /firmware/App/Modes/TreatmentStop.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged staging

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15788-pbo-ramps-up-to-400-mmhg-during-dialysate-circuit-priming' into develop

    • -10
    • +46
    /firmware/App/Controllers/PresOccl.c
DEN-15788: Added 10 sec venous pressure filter and publish venous pressure readings with 10 sec filter. Use 10 sec venous pressure filter for setting pressure windows after stabilization. Fix proposed UF rate and Tx duration after a UF volume change. Add simple pressure control to DPo speed during dialysate/dialyzer prime state to prevent pressure build up.

    • -10
    • +46
    /firmware/App/Controllers/PresOccl.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15660

DEN-15660 addressed a code review comment

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15628 changed the blood leak zero sequence counter from Z to G since Z is repeated twice in each zero sequence

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DEN-15660

DEN-15660, DEN-15721 added code to check the blood detection in treatment and service mode both

Merge branch 'staging' into DEN-15660-test_configuration_treatment_recovery

    • -7
    • +26
    /firmware/App/Modes/ModeTreatment.c
    • -3
    • +35
    /firmware/App/Services/Reservoirs.c