pretreatment

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
UI-BUG-S62: UI-BUG-S62-1

Addressed code reviews.

  1. … 5 more files in changeset.
DEN-11825: BP/HR screen has no instruction for user to read from bluetooth cuff or enter information

- Center aligned the notification

    • -0
    • +1
    ./connection/PreTreatmentConnectionStack.qml
  1. … 1 more file in changeset.
DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

Remainder items (summary):

- 3) Discuss whether to keep the manufacturing default on the slider since the user needs to confirm each parameter: Agreed to set to the minimum value

- 7) Arterial Pressure and Venous Pressure should be consistent with other parameters, the unit should be shown at the end: will be updated to have the units located in the title of the slider.

- Also fixed the title coloring when the slider has not been adjusted yet.

    • -5
    • +6
    ./connection/PreTreatmentConnectionStack.qml
  1. … 3 more files in changeset.
DEN-11825: BP/HR screen has no instruction

- Added the suggested on line instruction for the user.

    • -0
    • +8
    ./connection/PreTreatmentConnectionStack.qml
DEN-8145: Issues with UI Create Treatment screen

- Restricted the Patient ID entry to be max 20 characters and has to have at least 1 alphanumeric character without space, regarding the conversation with KeborMed.

  1. … 2 more files in changeset.
DEN-9704: Missing acknowledgement confirmation message/alarm when leaving BP/HR blank

- Added a script to later be used for the log export on the device, but it now only works on the VM since the device doesn't have the zip tool.

- Added a mimic function to simulate the BCuff measurement received and assigned F11 to it and it will only be enabled if the macro DEBUG_BCUFF_MIMIC is defined.

- Added opened and closed signals to the EntryDialog.qml to make it do the auto focus on the firstInput object when opened.

- Added setFocus() function in the BPHREntry.qml

- Removed the enabled option from doReset of the VTreatmentVitals

- And created two separate enableDialog for vital entry, and enable update for the auto update the properties.

- Added a constructor with systolic, diastolic, heartRate parameter to set the data on creation to be used in the message mimic.

- Added systolic, diastolic, heartRate parameters to the doTrigger of the VCommonAdjustmentVitals to open the vitals entry dialog with default values.

    • -2
    • +2
    ./connection/PreTreatmentConnectionStack.qml
  1. … 15 more files in changeset.
Code Cleanup

- DEN-11807: Treatment Confirmation - Rinseback Rate displays "min" instead of "mL/min"

- Renamed the Keyboard component name to VirtualKeyboard to resolve the InputPanel's Keyboard object/type conflict (There was no issue is was only a name conflict.)

- Preventing the Patient Id from being empty space string, and disabling the confirm in that case.

- Mode the create_update_folder.sh to the Scripts/local repository's folder.

- Added Qt logo and Diality Qt Colored log.

- Updated the alarmMapping.sh script to always checkout to common/staging and also print out the log on console.

-

  1. … 12 more files in changeset.
DEN-10614 [On Going Code Review] UI_DEV: Add Patient ID to Pre-Treatment Flow

- Pushed the code improvements

- Addressed the code review comments

    • -309
    • +197
    ./create/PreTreatmentCreate.qml
  1. … 8 more files in changeset.
DEN-10607: Screen

- Added the feature for the create treatment slider to be able to show OFF for zero values on label.

- Did a refactoring to clean up the SliderCreateTreatment.

  1. … 2 more files in changeset.
DEN-10607: Implementing Heparin

Completed

- Create Screen

- Confirm Screen

- MVC

- Log

- Dialin

  1. … 10 more files in changeset.
DEN-10607: Implementation [Screen]

- DEN-10608: Implementation [Navigation]

- DEN-10609: Implementation [Treatment Log]

    • -4
    • +54
    ./create/PreTreatmentCreateStack.qml
  1. … 8 more files in changeset.
DEN-9637: Update Design Documents

- Added the ability to enable/disable the vitals measurement logs [it's still unknown that how to stop in case of the treatment ended by alarm or system fault.]

- Added the Battery Level of the selected device.

    • -1
    • +1
    ./connection/PreTreatmentConnectionStack.qml
  1. … 18 more files in changeset.
DEN-9649: Implementation - Yocto configuration

- Addressed some frequent code review comments

- Fixed the UI POST AlarmTriggered message

- Added the deleteLater for the QQuickViewer pointer _viewer.

  1. … 113 more files in changeset.
DEN-8920 Implementation [Store Collected Data]

- Added MVC

- Added the TxLog and Log_Events and so...

- Added the timeout/interval

- ....

    • -3
    • +9
    ./connection/PreTreatmentConnectionStack.qml
  1. … 14 more files in changeset.
DEN-8918: BtC: Implementation [Screens] & DEN-8919: BtC: Implementation [Navigations]

- The view needs to be implemented in C++.

    • -0
    • +61
    ./PreTreatmentStack.qml
    • -3
    • +32
    ./connection/PreTreatmentConnectionStack.qml
  1. … 16 more files in changeset.
DEN-8490: Update the End Treatment Screens with the new Instruction base

- Also added the milliseconds to the logs

    • -2
    • +2
    ./connection/PreTreatmentUltrafiltration.qml
  1. … 35 more files in changeset.
DEN-8317: Implementation: Merge the Wi-Fi into the code.

- Mainly the WiFi integration

- Some other support tasks:

| Support : Instruction and Kishen images [ Calyton, Nick ]

| Support : Alarm Messaging Parser

| Support : Changed the Manager Screen into a Monitoring Dialog and can be shown by DblTap on SDCard Indication on top/right

| Support : Modified the Heat values to float on Monitoring Dialog

| Support/Bug : Fixed the Alarm Message Bug

  1. … 107 more files in changeset.
DEN-8317: Implementation: Merge the Wi-Fi into the code.

  1. … 11 more files in changeset.
DEN-8316: TxLog: Implementation [Simulator/Testing]

  1. … 53 more files in changeset.
DEN-8137: Confirm button in Confirm Treatment screen becomes disabled after enabled

- fixed the confirm button enabled.

  1. … 1 more file in changeset.
DEN-8145: Issues with UI Create Treatment screen

- Other than the comments bellow, we talked about the "--" was not correct which has been fixed.

1) The default value should not be shown on the slider that's not been configured yet

- Not sure that is has to be done but regardless, default values were not correct and fixed.

2) Dialyzer Temperature should show degree Celcius, not just C

- Dialyzer Temperature now has degree circle

3) Discuss whether to keep the manufacturing default on the slider since user needs to confirm each parameter

4) Concentrates and Dialyzer types should match with SRS /PRS

- Has been updated regarding the latest updates (BiCarb:PRS351, Dilyzer:PRS352)

5) Parameter value bounds back and forth on the slider during adjustment (display touch sensitivity issue)

6) Suggest to add + or - on the slider to provide ability to adjust precisely

7) Arterial Pressure and Venous Pressure should be consistent with other parameters, unit should be shown at the end

8) Suggest to display Arterial/Venous Pressure Limit as Arterial/Venous Low/High Pressure Limit and the current value as low/high mmHg

9) Suggest to display the current value of each parameter to the right of the slider

  1. … 7 more files in changeset.
DEN-8057: posTx: Implementation [Messaging] (HD StandBy subMode)

  1. … 16 more files in changeset.
DEN-7828: posTx: Implementation [Screens]

DEN-7830: posTx: Implementation [Navigation]

  1. … 11 more files in changeset.
DEN-7823: Meetings & Support - General purpose

- Added Alarm No Minimize so user can't minimize some alarms regarding the Alarm message received.

- Spell & doc correct

- Used MSG26 (Tx ranges) for the ultrafiltration pre-treatment max value.

- Fixed the Dynamic Instructions location on the device.

    • -1
    • +1
    ./connection/PreTreatmentUltrafiltration.qml
  1. … 11 more files in changeset.
DEN-7147: Code Review 0

- DEN-7823: Meetings & Support - General Purpose - Kushen Pre-Treatment Images.

  1. … 6 more files in changeset.
DEN-7832: Implementations [Saline and Heparin change request]

- Also removed the UF state from response.

  1. … 20 more files in changeset.
DEN-7147: Fixed a bug on arterial/venous pressures

- which was causing user to move low/high other than current default to be able to confirm for FW validation.

  1. … 2 more files in changeset.
DEN-7147: Fixed a bug on Sub-StepIndicator on bottom of the instructions

- Which was causing that to not to show correct number of the steps and titles.

- removed #include "MModel.h" in all the possible cases and include the specific model since the MModel has too much include.

- Showing the Heparin setup in the CART_INSTALL state only if we have a heparin volume setup.

- created a common template function name enumString in Gui namespace to get the enum name as string.

  1. … 71 more files in changeset.