hdfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Typo.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Blood pump only stops for user stop.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Remove redundant condition.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Simplified. Addressed CR comments.

    • -44
    • +21
    /firmware/App/Modes/TreatmentRecirc.c
DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Add restart flag.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Control update for Christina's observation

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Updates to address comments.

    • -30
    • +35
    /firmware/App/Modes/TreatmentRecirc.c
DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Corrected comments and alignment

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

HW update only on flag changes

    • -12
    • +21
    /firmware/App/Modes/TreatmentRecirc.c
DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Minimal change to do "all stop" for user stop button in treatment recirculation.

    • -0
    • +65
    /firmware/App/Modes/TreatmentRecirc.c
DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Reversed working revisions, to try minimal change.

    • -6
    • +3
    /firmware/App/Controllers/DialInFlow.c
    • -6
    • +3
    /firmware/App/Controllers/DialOutFlow.c
    • -45
    • +16
    /firmware/App/Modes/TreatmentRecirc.c
DEN-15741_Dpi Does Not Stop When Recirculation IS Paused 2

Update to user stop state, and hard stop functions.

    • -3
    • +6
    /firmware/App/Controllers/DialInFlow.c
    • -3
    • +6
    /firmware/App/Controllers/DialOutFlow.c
    • -16
    • +45
    /firmware/App/Modes/TreatmentRecirc.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge remote-tracking branch 'origin/DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-2' into staging

Merge remote-tracking branch 'origin/DEN-15893-cartridge-lever-is-open-alarm-when-cartridge-lever-is-closed' into staging

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15804

DEN-15804 reverted the saline bolus changes

DEN-15893-cartridge-lever-is-open-alarm-when-cartridge-lever-is-closed

cleared alarm 120 in self test

DEN-15804 changed the blood flow coefficients

DEN-15893-cartridge-lever-is-open-alarm-when-cartridge-lever-is-closed

Cleared alarm 120, prior to initiating alarm 85

Required latch closed before allowing alarm 85

Paused in WET self test until bubbles clear or latch closed.

DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-2

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge remote-tracking branch 'origin/DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus' into staging