Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    ./App/Controllers/ConductivitySensors.c
    • -2
    • +2
    ./App/Controllers/TemperatureSensors.c
Merge branch 'DEN-15702-alarm-150-inlet-water-conductivity-high-does-not-disappear-on-standby-screen-when' into staging

    • -45
    • +59
    ./App/Controllers/TemperatureSensors.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./App/Controllers/TemperatureSensors.c
    • -1
    • +1
    ./App/Modes/ModeChemicalDisinfectFlush.c
    • -1
    • +1
    ./App/Modes/ModeHeatDisinfectActiveCool.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15702 Code review update - remove spaces.

    • -1
    • +1
    ./App/Controllers/ConductivitySensors.c
    • -1
    • +1
    ./App/Controllers/TemperatureSensors.c
DEN-15702 Indent for code review comments.

    • -57
    • +57
    ./App/Controllers/ConductivitySensors.c
    • -49
    • +49
    ./App/Controllers/TemperatureSensors.c
DEN-15608 Correct monitoring of voltages

Merge branch 'staging' into DEN-15660-test_configuration_treatment_recovery

    • -22
    • +54
    ./App/Controllers/ConcentratePumps.c
    • -2
    • +9
    ./App/Modes/ModeHeatDisinfectActiveCool.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15660 addressed the code review comments

    • -1
    • +1
    ./App/Controllers/TemperatureSensors.c
DEN-15660 addressed the code review comments

    • -5
    • +5
    ./App/Controllers/TemperatureSensors.c
    • -21
    • +24
    ./App/Modes/ModeChemicalDisinfect.c
DEN-15702 Check VPI Open for Inlet Conductivity, Pressure, temperature. If VPI Closed, clear alarms.

    • -8
    • +17
    ./App/Controllers/ConductivitySensors.c
    • -10
    • +24
    ./App/Controllers/TemperatureSensors.c
DEN-15809 added code to clear the dialysate generation alarm 147

DEN-15809 updated the code to trigger another alarm if the empty bottle alarms have been triggered.:

DEN-15660, DEN-15781, added code to not check the dialysate temperature sensorswhen the dialysate flow is zero. DEN-15808, Added code to process the empty jug alarm

    • -6
    • +10
    ./App/Controllers/TemperatureSensors.c
DEN-15660, DEN-15767, DEN-15781 cleaned up the SystemCommMessages.c for messages list script. Added moving average for TDi and TRo sensors for dialysate temperature drift check

    • -4
    • +81
    ./App/Controllers/TemperatureSensors.c
DEN-15802 Correct alarm 191.192 testing based upon specific modes.

STRL-15802 Add Heat Disinfect Active Cool to logic to test for 191, 192

DEN-15715 added a command to close VRD1 prior to transitioning to chemical disinfect cancel water path

    • -11
    • +11
    ./App/Modes/ModeChemicalDisinfect.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15691-heat-disinfect-reservoir-leak-alarm-triggers-after-5s-when-srsdg-512-and-521-say-it

Corrected comment alignment

DEN-15691-heat-disinfect-reservoir-leak-alarm-triggers-after-5s-when-srsdg-512-and-521-say-it

Undated 5second timeout to 10 seconds

DEN-15697 changed the code to correctly set the drain timeout in active cool mode

    • -9
    • +8
    ./App/Modes/ModeHeatDisinfectActiveCool.c
DEN-15619 removed the variable to not trigger a temperature not reached to target alarm in chemical disinfect