Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DENBUG-163: High Venous Pressure Alarm 130 Triggers During Normal Treatment Operation

Added air trap raw lower level and air trap raw upper level to be logged and displayed on the diagnostics screen

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -3
    • +3
    ./adjustment/common/VCommonAdjustmentVitals.cpp
    • -1
    • +1
    ./adjustment/posttreatment/VPostTreatmentAdjustDisposablesRemovalConfirm.cpp
    • -1
    • +1
    ./adjustment/posttreatment/VPostTreatmentAdjustDisposablesRemovalConfirm.h
    • -1
    • +1
    ./adjustment/posttreatment/VPostTreatmentAdjustPatientDisconnectionConfirm.cpp
    • -1
    • +1
    ./adjustment/posttreatment/VPostTreatmentAdjustPatientDisconnectionConfirm.h
    • -1
    • +1
    ./adjustment/pretreatment/VPreTreatmentAdjustmentConsumablesConfirm.cpp
    • -1
    • +1
    ./adjustment/pretreatment/VPreTreatmentAdjustmentConsumablesConfirm.h
    • -1
    • +1
    ./adjustment/pretreatment/VPreTreatmentAdjustmentDisposablesConfirm.cpp
    • -1
    • +1
    ./adjustment/pretreatment/VPreTreatmentAdjustmentDisposablesConfirm.h
    • -1
    • +1
    ./adjustment/pretreatment/VPreTreatmentAdjustmentDisposablesPrime.cpp
  1. … 598 more files in changeset.
DENBUG-204 updated the message payload of message 0x0B00 by adding an integer at the end of the structure

  1. … 5 more files in changeset.
DENBUG-182: Phantom Alarm Screen Appears After Clearing Last Alarm

- the order of the QML PROPERTYs cannot exactly predicted therefore setting them if not needed has to be avoided.

- set the alarmID to 0 to trigger the next change since we skip the property updated if the alarmID is 0.

  1. … 2 more files in changeset.
DENBUG-136: Main treatment screen fails to prompt user with vital measurements entry after set interval

- the issue cannot be reproduced.

- added a count down timer for the vital entry to better understand how and when the timer works on each interval.

    • -0
    • +2
    ./adjustment/common/VCommonAdjustmentVitals.cpp
  1. … 1 more file in changeset.
CS-3: CS: Heparin OFF o the HCP portal Treatment report

- Fixed some issues in the create treatment screen on heparin type

- Combined the separated values for log and GUI into one values for both.

- Added the functionality for the TreatmentConfirm (Pre-Tx), TreatmentReview (Ppost-Tx), TreatmentReport (log) for the heparin values.

  1. … 7 more files in changeset.
CS-5: UI: TxCode generation on device

  1. … 9 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
DEN-16286: UI: CS: In the Treatment log map the Alarm/Event ID to the description

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 55 more files in changeset.
DEN-16256: Updated - UI.P1B : Display UF Rate

Added QML text to display the rate

Include the 11th parameter

  1. … 3 more files in changeset.
DEN-16256: Added - UI.P1B : Display UF Rate

Add the variable to the 0x0B00 message model to include UF Current set rate

Setting of the value received from HD for UF Current Set Rate

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./adjustment/posttreatment/VPostTreatmentAdjustPatientDisconnectionConfirm.cpp
    • -1
    • +1
    ./adjustment/posttreatment/VPostTreatmentAdjustPatientDisconnectionConfirm.h
  1. … 17 more files in changeset.
DEN-14815: Cartridge Removal Screen is Displayed Prematurely

- Added the Response model and updated the view for MAdjustPatientDisconnectionConfirmResponse

    • -2
    • +18
    ./adjustment/posttreatment/VPostTreatmentAdjustPatientDisconnectionConfirm.cpp
    • -1
    • +2
    ./adjustment/posttreatment/VPostTreatmentAdjustPatientDisconnectionConfirm.h
  1. … 8 more files in changeset.
DEN-14815: Cartridge Removal Screen is Displayed Prematurely

- Added the message ID_AdjustPatientDisconnectionConfirmReq = 0x0600, // 6

    • -2
    • +12
    ./adjustment/posttreatment/VPostTreatmentAdjustPatientDisconnectionConfirm.cpp
    • -1
    • +3
    ./adjustment/posttreatment/VPostTreatmentAdjustPatientDisconnectionConfirm.h
  1. … 8 more files in changeset.
DEN-14815: Cartridge Removal Screen is Displayed Prematurely

- Modified the Patient Disconnection confirm to notify since there is no response and UI just notifies the HD.

- This way the actual confirm which is on the Treatment Review screen Next button can have better name.

- Also modified the ID_AdjustDisposablesRemovalConfirmRsp enum to match its req.

    • -2
    • +2
    ./adjustment/posttreatment/VPostTreatmentAdjustPatientDisconnectionConfirm.cpp
    • -1
    • +1
    ./adjustment/posttreatment/VPostTreatmentAdjustPatientDisconnectionConfirm.h
  1. … 7 more files in changeset.
DEN-15911: Post-treatment screen displays heparin type and heparin concentration even if heparin rate and dispensing were disabled in pre-treatment

- Separated the Log and the Gui values for now to address the cloud log later, when the cloud is ready to not to reject Logs to have NONE.

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 13 more files in changeset.
DEN-15487: Updated - UI: POST: UI needs to report an error if the configuration files, or parameter(s) are missing.

DEN-16104: UI: Reorder the sequence of the messages in POST from UI to HD

  1. … 12 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 15 more files in changeset.
Merge branch 'DEN-15905-UI-BN-S102-Log178' into staging

  1. … 1 more file in changeset.
  1. … 2 more files in changeset.
DEN-15514: UI: In-Progress Indicator Missing During Cartridge Priming Steps

- Added some possible solution and will see what is the feedback.

  1. … 2 more files in changeset.
DEN-15487: Updated - UI: POST: UI needs to report an error if the configuration files, or parameter(s) are missing.

Updated with changes per CR request

  1. … 4 more files in changeset.
DEN-15487: Added - UI: POST: UI needs to report an error if the configuration files, or parameter(s) are missing.

- the checking of POST for settings CRC

- error handling when reading the settings file

- error handling when configuration parameters are incorrect

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
DEN-15491: Updated - Usability: UI: alarm list unique field for Alarm Descriptor

Renamed the alarm name variable to be listTitle per CR request

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    ./data/treatment/VHDTreatmentStatesData.h
  1. … 7 more files in changeset.
DEN-15769: UI: Usability: Remove Option to increase/decrease treatment time after UF volume adjustments

- Fixed the added new state from common develop (blood recover)

    • -0
    • +1
    ./data/treatment/VHDTreatmentStatesData.h
DEN-14934: Rinseback "Back to Treatment" Workflow Missing Patient Reconnection Instructions (Phase 1)

    • -0
    • +1
    ./data/treatment/VHDTreatmentStatesData.h
  1. … 6 more files in changeset.