MessageGlobals.h

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 31 more files in changeset.
DEN-16864: UI Bloodflow message is incomplete

- Updated the payload of the following messages: 0x002C, 0x0005, 0x0008, 0x006C.

  1. … 16 more files in changeset.
DEN-16862: Add a command to reset HD during UI software install

- Added a command to request an HD reset during the UI software and while HD is in service mode.

  1. … 6 more files in changeset.
DEN-16204: UI: FW: Pure Water mode turning on intermittently. - Added a new payload to send the current firmware value with the response to a set from UI - Changed the RO only mode code to not flip the UI RO only status if the firmware rejected the set value and just get the status in the firmware

  1. … 4 more files in changeset.
DEN-16851: New Alarm for Increased Occlusion Sensitivity - Added the following fields to mssage 0x0009: arterial long filter, venous long filter, blood pump occl long filter, partial occl base line

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
DEN-16773: Add Chemical Disinfect Enable To the Institutional Record

Added enable chemical disinfect to the institutional configurations

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 28 more files in changeset.
DENBUG-199: Alarm 280 (DG Fault: DRP Direction Fault) Triggered during Pre-Treatment Sequence

Added the Maxon RPM to the handled messages to be logged.

  1. … 6 more files in changeset.
DEN-16693: Institutional Settings Implementation

Initial commit of the inititutional record MVC implementation.

  1. … 21 more files in changeset.
DENBUG-163: High Venous Pressure Alarm 130 Triggers During Normal Treatment Operation

Added air trap raw lower level and air trap raw upper level to be logged and displayed on the diagnostics screen

  1. … 8 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
DEN-16444: UI: UI Logs do not Capture TDi and TRo Moving Average Values

Updated the application to incorporate the two new float values in the DG temperature sensors payload.

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 612 more files in changeset.
DENBUG-204 updated the message payload of message 0x0B00 by adding an integer at the end of the structure

  1. … 6 more files in changeset.
DEN-16531: UI: The reservoir message structure has changed on FW and UI needs to adjust.

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 56 more files in changeset.
Merge branch 'DEN-16335-UI-VD-S105_HD_DG_UsageCounterDisplay' into staging

  1. … 3 more files in changeset.
DEN-16262: Added - UI.P1B : HD/DG usage counter

Added the MV for the HD usage information

Modify the TouchGrid to accomodate indentation of subCategories

  1. … 16 more files in changeset.
DEN-16256: Updated - UI.P1B : Display UF Rate

Added QML text to display the rate

Include the 11th parameter

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 20 more files in changeset.
DEN-14815: Cartridge Removal Screen is Displayed Prematurely

- Added the Response model and updated the view for MAdjustPatientDisconnectionConfirmResponse

  1. … 9 more files in changeset.
DEN-14815: Cartridge Removal Screen is Displayed Prematurely

- Added the message ID_AdjustPatientDisconnectionConfirmReq = 0x0600, // 6

  1. … 9 more files in changeset.
DEN-14815: Cartridge Removal Screen is Displayed Prematurely

- Modified the Patient Disconnection confirm to notify since there is no response and UI just notifies the HD.

- This way the actual confirm which is on the Treatment Review screen Next button can have better name.

- Also modified the ID_AdjustDisposablesRemovalConfirmRsp enum to match its req.

  1. … 8 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
DEN-14961: Updated - The UI RTC needs to get synced with HD RTC

Renamed the class and objects to be HDRTC* per CR request

  1. … 12 more files in changeset.
DEN-14961: Added - The UI RTC needs to get synced with HD RTC

Add the HD-UI RTC message model to the UI to do a single HD-UI RTC sync

  1. … 11 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 54 more files in changeset.
DEN-15359: Usability: Arterial and Venous Pressure Alarm Limits

- DEN-15387: UI: Updates for treatment parameter entry (A/V alarm window)

- DEN-15388: UI: Updates for in-treatment adjustment of alarm parameter window (pressure alarm screen)

- Monitoring components for the TxHome, Rinseback, BloodPrime

  1. … 35 more files in changeset.
DEN-15316: Usability: UI: Remove Extra Button for Disinfection

- Implemented the MVC and messaging.

  1. … 18 more files in changeset.