DENBUG-294 addressed the code review comment

DENBUG-294 addressed the code review comments

    • -16
    • +15
    /firmware/App/Modes/ModeInitPOST.c
DENBUG-294 added code to send the serial number and institutional record in POST

    • -15
    • +49
    /firmware/App/Modes/ModeInitPOST.c
DENBUG-288 revered the alarm management code

DENBUG-290 fixed the blood leak code

DENBUG-290, DEN-172020 added code to not trigger the blood leak alarm in saline bolus

DENBUG-288 added to code in alarm management to special handle alarm 129

DENBUG-292 changed the default rinseback volume to 300 mL

DEN-17161 changed calcium and postassium data types from U32 to F32 in treatment logs. DENBUG-290 added code to not trigger alarms 126 and 129 in treatment recirculation

DENBUG-290 updated the blood leak driver to not check for blood recover until the dialysate inlet pump is not running. Changed the rinseback target volume from 350 mL to 400 mL

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEBUG-285

DENBUG-285 updated the VectorCAST test cases

DENBUG-285 changed the target blood flow rate at treatment end to 150 mL/min

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-17156

DEN-17165 updated the code to make the blood prime set a signed integer to be able to calculate the ramp down. Changed the control to be able to do the ramp up and ramp down

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DENBUG-285

DENBUG-285 updated the VectorCAST test cases

DENBUG-285 addressed the code review comment and updated the rinseback additional to 250 mL/min

DENBUG-285 updated the target blood flow rates in saline bolus and treatment end

DENBUG-285 fixed the VBT in air trap

DENBUG-285 removed the VBV code in air trap

    • -18
    • +0
    /firmware/App/Controllers/AirTrap.c
Merge branch 'DENBUG-285-alarm-141-air-trap-fill-timeout' of bitbucket.org:diality-cloud/hdfirmware into DENBUG-285-alarm-141-air-trap-fill-timeout

DENBUG-285 updated the code per the latest requests: 1. Reverted the VBV in airtrap. 2. Changed blood prime to 250mlPerMin and 3. Changed the min venous pressure to 0 mmHg

    • -14
    • +10
    /firmware/App/Modes/BloodPrime.c
DENBUG-285 added code to set the VBV to close in air trap open state