Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 17 more files in changeset.
DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Modified message 0xBA to send up to 4 float payloads to the UI

- Changed MDuetConfirmHD to receive and process the new payloads

- Changed VConfirm to update the confirm messages that have {} in them and replace them with actual values

  1. … 6 more files in changeset.
StagingMerge - DEN-17260-ui-dn-release-21 (13 commits squashed)

DEN-17260: UI-DN-Release-21

- DEN-17261: Corrected the treatment logs doxygen and data type for potassium and calcium.

DEN-17260: UI-DN-Release-21

- DENBUG-298: Removed the Maximum Saline Bolus State

- DENBUG-298: Added Deliver Saline Bolus in treatment stop.

DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating

- Reduced the font size of fluid value font from 32 to 28.

DEN-17269: Change_version_format_from_dot_to_dash

- Changed the format of the version from dot to dash before compatibility value.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Modified message 0xBA to send up to 4 float payloads to the UI

- Changed MDuetConfirmHD to receive and process the new payloads

- Changed VConfirm to update the confirm messages that have {} in them and replace them with actual values

DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating

- Updated the TreatmentSaline.qml file to clear the reject reasons associated with saline bolus once the treatment state has changed.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Updated the VConfirm to trigger the confirm messages even if the ID has not changed.

DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating

- Updated the code to clear the saline bolus related reject reasons upon changing the saline bolus states.

DENBUG-298: User unable to deliver saline bolus during alarm conditions where the blood pump is recirculating

- Added code to clear the saline bolus reject reason upon changing the treatment states.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Updated the code to round up the UF rate that is displayed on the UF rate change screen.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Updated the current UF rate in the VTreatmentUltrafiltration to be rounded up.

- Removed commented code.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Made UF Rate related values in VTreatmentAdjustmentUltrafiltrationEdit to be rounded up prior to displaying on the UI.

DENBUG-302: When treatment time is updated mid-treatment, update UF rate to keep the volume prescription the same

- Addressed the code review comments

  1. … 17 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 619 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 619 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
DEN-17093: Blood Leak Zeroing in Treatment

- Added a new enum in dialysis for blood leak zeroing

- Updated the payloads for message 0x6C

- Changed "Date And Time" to "Date and Time"

DEN-17093: Blood Leak Zeroing in Treatment

- Added a new field to message 0x2F00 to log dialysate pump inlet to blood leak detector flush volume

- Removed the blood leak zeroing state enum read

DEN-17093: Blood Leak Zeroing in Treatment

- Fixed a typo

DEN-17093: Blood Leak Zeroing in Treatment

- Updated the payload of message 0x2F

- Updated the payload of message 0x0F

- Updated the payload of message 0x6C

DEN-17093: Blood Leak Zeroing in Treatment

- Fixed blood leak view class

  1. … 10 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
DEN-17093: Blood Leak Zeroing in Treatment

- Updated the payload of message 0x2F

- Updated the payload of message 0x0F

- Updated the payload of message 0x6C

  1. … 6 more files in changeset.
DEN-17093: Blood Leak Zeroing in Treatment

- Added a new field to message 0x2F00 to log dialysate pump inlet to blood leak detector flush volume

- Removed the blood leak zeroing state enum read

  1. … 4 more files in changeset.
DEN-17093: Blood Leak Zeroing in Treatment

- Added a new enum in dialysis for blood leak zeroing

- Updated the payloads for message 0x6C

- Changed "Date And Time" to "Date and Time"

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 16 more files in changeset.
Release 17 Code Review branch

git checkout release/stagingmerge-DEN-17017-ui-bn-release-17

git diff HEAD staging > ../stagingmerge-DEN-17017-ui-bn-release-17.patch

git apply ../stagingmerge-DEN-17017-ui-bn-release-17.patch

  1. … 16 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
DEN-16989: Treatment Information Not Showing on Information Screen

- Updated the UI code to just show the total treatment hours and the hours since last service in hours.

- Updated the UI code to just to show the last treatment start date in YYYY/MM/DD.

- Removed the last reset date and the CRC,

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 31 more files in changeset.
DEN-16851: New Alarm for Increased Occlusion Sensitivity - Added the following fields to mssage 0x0009: arterial long filter, venous long filter, blood pump occl long filter, partial occl base line

DEN-16204: UI: FW: Pure Water mode turning on intermittently.

- Added a new payload to send the current firmware value with the response to a set from UI

- Changed the RO only mode code to not flip the UI RO only status if the firmware rejected the set value and just get the status in the firmware

DEN-16862: Add a command to reset HD during UI software install

- Added a command to request an HD reset during the UI software and while HD is in service mode.

DEN-16864: UI Bloodflow message is incomplete

- Updated the payload of the following messages: 0x002C, 0x0005, 0x0008, 0x006C.

DEN-16204: UI: FW: Pure Water mode turning on intermittently

- Changed the order of the UI response message from firmware so the accept is always the first field.

DEN-16204: UI: FW: Pure Water mode turning on intermittently

- Addressed the code review comments.

  1. … 30 more files in changeset.
DEN-16872 Blood priming screen appeared during treatment

- reverted back a residual modification which caused to break the ACK

DEN-16872 Blood priming screen appeared during treatment

- added caching to the UI application logs.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 31 more files in changeset.
DEN-16864: UI Bloodflow message is incomplete

- Updated the payload of the following messages: 0x002C, 0x0005, 0x0008, 0x006C.

  1. … 16 more files in changeset.
DEN-16862: Add a command to reset HD during UI software install

- Added a command to request an HD reset during the UI software and while HD is in service mode.

  1. … 6 more files in changeset.
DEN-16204: UI: FW: Pure Water mode turning on intermittently. - Added a new payload to send the current firmware value with the response to a set from UI - Changed the RO only mode code to not flip the UI RO only status if the firmware rejected the set value and just get the status in the firmware

  1. … 4 more files in changeset.
DEN-16851: New Alarm for Increased Occlusion Sensitivity - Added the following fields to mssage 0x0009: arterial long filter, venous long filter, blood pump occl long filter, partial occl base line

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
DEN-16773: Add Chemical Disinfect Enable To the Institutional Record

Added enable chemical disinfect to the institutional configurations

DEN-16773: Add Chemical Disinfect Enable To the Institutional Record

- Added code to display the reject reason in chemical disinfect flush

DEN-16773: Add Chemical Disinfect Enable To the Institutional Record

- Addressed the code review comment

DEN-16773: Add Chemical Disinfect Enable To the Institutional Record

- Addressed a code review comment

  1. … 6 more files in changeset.
DEN-16773: Add Chemical Disinfect Enable To the Institutional Record

Added enable chemical disinfect to the institutional configurations

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 28 more files in changeset.
DENBUG-163: High Venous Pressure Alarm 130 Triggers During Normal Treatment Operation

Added air trap raw lower level and air trap raw upper level to be logged and displayed on the diagnostics screen

DEN-16693: Institutional Settings Implementation

Initial commit of the inititutional record MVC implementation.

DEN-16693: Institutional Settings Implementation

Updated the qml file to display the insititutional record. Updated the model and view classes.

Bamboo Commit: Updated MAlarmMapping.cpp

DENBUG-199: Alarm 280 (DG Fault: DRP Direction Fault) Triggered during Pre-Treatment Sequence

Added the Maxon RPM to the handled messages to be logged.

DEN-16693: Institutional Settings Implementation

Updated the institutional record screen columns.

DEN-16693: Institutional Settings Implementation

Updated the units width in the institutional record.

DEN-16693: Institutional Settings Implementation

Changed the prescribed UF volume to L from mL.

DEN-16693: Institutional Settings Implementation

Updated the title of the Institutional Configurations and updated the heparin units

DEN-16693: Institutional Settings Implementation

Changed the name of the Institutional Configurations to Institutional Configs to be able to fit in the tab

DEN-16693: Institutional Settings Implementation

- Touch grid column spacing shrank from 100 to 50.

- Settings column width expanded from 300 to 325.

- Changed the Institutional Configs to Institutional Configurations on settings items.

DEN-16693: Institutional Settings Implementation

- Addressed the code review comments

  1. … 34 more files in changeset.