DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- Reverted back the Heparin min to 0.0 due to the slider's issue which disables the OFF feature, since the value 0.0 becomes an invalid.

- A minor update on the POST progress display to start when 0 to max of complete enum -1.

- An idea of fixing the slider touch issue. [Is DISABLED for now, for more test]

  1. … 11 more files in changeset.
RESOLVED.

RESOLVED.

Disregard. Found where it is used.

Disregard. Found where it is used.

What is the purpose of this alias?

What is the purpose of this alias?

Why commented out?

Why commented out?

Why commented out?

Why commented out?

Why is this commented out?

Why is this commented out?

When will we do this?

When will we do this?

File name does not appear to match image.

File name does not appear to match image.

The declaration in the header file has const for all 3 parameters.

The declaration in the header file has const for all 3 parameters.

That is a very good point. No one ever mentioned that. Let me investigate, why I had all these repeated else. Could be for performance not to check next, but there is another way for that too. If ...

That is a very good point.
No one ever mentioned that.

Let me investigate, why I had all these repeated else.
Could be for performance not to check next, but there is another way for that too.
If I can't find a quick answer for that let me create a case for that since it is everywhere.

RESOLVED.

RESOLVED.

There has to be and there is. It is just a long line and those little ones are at the end.

There has to be and there is.
It is just a long line and those little ones are at the end.

It seems you have multiple else statements that do the same, is that right? If yes, why multiple elses?

It seems you have multiple else statements that do the same, is that right? If yes, why multiple elses?

Isn't is safer to have a break at the end of each case?

Isn't is safer to have a break at the end of each case?

DEN-11814: Heparin Minimum Dispensing Parameters Should Be 0.2

- updated the min range regarding the recent PRS-347, PRS-348

- Fixed the minimum precision of the heparin.

  1. … 4 more files in changeset.
DEN-8145: Issues with UI Create Treatment screen

Remainder items (summary):

- 3) Discuss whether to keep the manufacturing default on the slider since the user needs to confirm each parameter: Agreed to set to the minimum value

- 7) Arterial Pressure and Venous Pressure should be consistent with other parameters, the unit should be shown at the end: will be updated to have the units located in the title of the slider.

- Also fixed the title coloring when the slider has not been adjusted yet.

  1. … 4 more files in changeset.
UI-BUG-S62
UI-BUG-S62
DEN-9704: Missing acknowledgement confirmation message/alarm when leaving BP/HR blank

- Added a script to later be used for the log export on the device, but it now only works on the VM since the device doesn't have the zip tool.

- Added a mimic function to simulate the BCuff measurement received and assigned F11 to it and it will only be enabled if the macro DEBUG_BCUFF_MIMIC is defined.

- Added opened and closed signals to the EntryDialog.qml to make it do the auto focus on the firstInput object when opened.

- Added setFocus() function in the BPHREntry.qml

- Removed the enabled option from doReset of the VTreatmentVitals

- And created two separate enableDialog for vital entry, and enable update for the auto update the properties.

- Added a constructor with systolic, diastolic, heartRate parameter to set the data on creation to be used in the message mimic.

- Added systolic, diastolic, heartRate parameters to the doTrigger of the VCommonAdjustmentVitals to open the vitals entry dialog with default values.

  1. … 15 more files in changeset.
RESOLVED.

RESOLVED.

RESOLVED

RESOLVED

Fixed. Thanks for your attention to the details.

Fixed.
Thanks for your attention to the details.

Some functions and variables in this file use PatientID instead of PtientID, was this spelling intentional?

Some functions and variables in this file use PatientID instead of PtientID, was this spelling intentional?

Added

Added

Needs function header?

Needs function header?

RESOLVED