DEN-16444: UI: UI Logs do not Capture TDi and TRo Moving Average Values

Updated the application to incorporate the two new float values in the DG temperature sensors payload.

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 608 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DEN-16637, DEN-16640, DEN-16641 : P1_Staging Dry-Run found issues.

DEN-16637: SW: Top Alarm on Debug Screen Does Not Change When Alarm is Cleared

DEN-16640: SW: Blood Measurement Interval Set to OFF During Treatment When Set to 5 min in Pre-treatment

DEN-16641: UI: Heparin Syringe Installation Screen Missing

  1. … 4 more files in changeset.
DEN-16637, DEN-16640, DEN-16641 : P1_Staging Dry-Run found issues.

DEN-16637: SW: Top Alarm on Debug Screen Does Not Change When Alarm is Cleared

DEN-16640: SW: Blood Measurement Interval Set to OFF During Treatment When Set to 5 min in Pre-treatment

DEN-16641: UI: Heparin Syringe Installation Screen Missing

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 608 more files in changeset.
DENBUG-204 updated the message payload of message 0x0B00 by adding an integer at the end of the structure

DENBUG-204 updated the diagnostics screens qml with the latest dialysate outlet pump broadcast changes

DENBUG-204 aligned the commented out code in the diagnostics qml

  1. … 7 more files in changeset.
DENBUG-204 aligned the commented out code in the diagnostics qml

DENBUG-204 updated the diagnostics screens qml with the latest dialysate outlet pump broadcast changes

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 56 more files in changeset.
DEN-16258: Updated - UI.P1B : Bluetooth status indicator

Modify to include vendor name and model name in status indicator

Change the data that is displayed in the status dialog

Status dialog now handles single line text

  1. … 5 more files in changeset.
DEN-16257: Updated - UI.P1B : WiFi status indicator

Add security types and signal strength level to the status dialog

Modify wifi status dialog to have space between icon and dialog body

Changed look and feel of status dialog based on review call

  1. … 5 more files in changeset.
DEN-16251: PRS 538 Not Met : Able to Perform Factory Reset With One Confirmation

  1. … 6 more files in changeset.
DEN-16258: Updated - UI.P1B : Bluetooth status indicator

Include automatic height setting based on rows of information

Make use of the displayInformation bool

  1. … 1 more file in changeset.
DEN-16258: Added - UI.P1B : Bluetooth status indicator

The bluetooth indicator QML component

A re-usable StatusDialog component that opens a small dialog to display info passed

Remove the old text bluetooth debug string

  1. … 9 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Alarm Dialog font sizes have adjusted to meet the requirement

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 15 more files in changeset.
  1. … 1 more file in changeset.
DEN-15078 DRP Cur and HDrst Values Display Too Many Float Numbers

  1. … 2 more files in changeset.
Release feedback: Drain Pump currentA parameter precision

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 57 more files in changeset.
Merge branch 'DEN-14936-UI-BN-S93' into staging

  1. … 2 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

  1. … 10 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

DEN-14853 UI Out Of Sync With Firmware On Handled Messages

  1. … 8 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

- MSG_ID 0x0B

- Added the keepAlive/UI-CheckIn to the list of AckRequired so if HD won't ack UI, the Alarm, "HD Communication Failure" will be displayed.

  1. … 6 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

  1. … 19 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 14 more files in changeset.
DEN-13152 : Flow Rate Settings Display "--" or "0" When Treatment is Paused

- Messages 0x0500 and 0x0800 has been updated and the Prescribed Flow Rate is being used.

- The diagnostics screen updated.

- The treatment main screen updated.

- The treatment flow adjustment screen updated.

- the Flows on the commit has 10f39b63ce8a992280ab123f05a5eb44a27c121b has been separated which has to be reverted back since these two messages almost all the time have the same payloads.

  1. … 11 more files in changeset.