pages

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-16002: Treatment Logs Do Not Display on UI

DEN-16213: UI: Incorrect status message upon entering Factory Reset/Decommission screen

    • -1
    • +4
    ./settings/SettingsServicePassword.qml
DEN-16284: UI: Cybersecurity: Disable the SSH access by default

    • -48
    • +132
    ./settings/SettingsRootSSHAccess.qml
  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./posttreatment/PostTreatmentStack.qml
    • -1
    • +1
    ./pretreatment/PreTreatmentDisposablesStack.qml
    • -1
    • +1
    ./pretreatment/create/PreTreatmentCreateStack.qml
  1. … 17 more files in changeset.
DEN-14815: Cartridge Removal Screen is Displayed Prematurely

- Added the QML code.

    • -1
    • +12
    ./posttreatment/PostTreatmentStack.qml
DEN-14815: Cartridge Removal Screen is Displayed Prematurely

- Added the message ID_AdjustPatientDisconnectionConfirmReq = 0x0600, // 6

    • -2
    • +3
    ./posttreatment/PostTreatmentStack.qml
  1. … 9 more files in changeset.
DEN-15911: Post-treatment screen displays heparin type and heparin concentration even if heparin rate and dispensing were disabled in pre-treatment

- Separated the Log and the Gui values for now to address the cloud log later, when the cloud is ready to not to reject Logs to have NONE.

    • -1
    • +2
    ./posttreatment/PostTreatmentStack.qml
    • -1
    • +1
    ./pretreatment/PreTreatmentDisposablesStack.qml
    • -1
    • +4
    ./pretreatment/create/PreTreatmentCreateStack.qml
  1. … 5 more files in changeset.
DEN-15112: Added - UI: UI Does not Display (Pending) Treatment Logs on Export Logs Screen

Add the log list model for SD card and USB in CPP

Updated the SettingLogExport QML to reflect CPP model addition

    • -122
    • +28
    ./settings/SettingsExportLogs.qml
  1. … 8 more files in changeset.
DEN-15112: Updated - UI: UI Does not Display (Pending) Treatment Logs on Export Logs Screen

Use the correct path array to also find the treatment logs in the encrypted partition

  1. … 2 more files in changeset.
DEN-15112: Updated - UI: UI Does not Display (Pending) Treatment Logs on Export Logs Screen

Add doxygen comments and debug comments

  1. … 3 more files in changeset.
DEN-15112: Updated - UI: UI Does not Display (Pending) Treatment Logs on Export Logs Screen

Add improved USB insert/remove handling that does not rely on polling

Add updates to models base on directory changes

    • -101
    • +18
    ./settings/SettingsExportLogs.qml
  1. … 17 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    ./pretreatment/PreTreatmentPrimeStack.qml
  1. … 16 more files in changeset.
    • -1
    • +1
    ./pretreatment/PreTreatmentPrimeStack.qml
  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    ./treatment/adjustments/TreatmentAdjustmentPressuresLimits.qml
  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 14 more files in changeset.
DEN-15514: UI: In-Progress Indicator Missing During Cartridge Priming Steps

- Added some possible solution and will see what is the feedback.

    • -6
    • +25
    ./pretreatment/PreTreatmentPrimeStack.qml
  1. … 3 more files in changeset.
DEN-15112: Added - UI: UI Does not Display (Pending) Treatment Logs on Export Logs Screen

Updated the filemodel to have a proxyModel for filtering and sorting actions

    • -42
    • +13
    ./settings/SettingsExportLogs.qml
  1. … 13 more files in changeset.
DEN-15514: UI: In-Progress Indicator Missing During Cartridge Priming Steps

- The issue on the FW side which not updating the corresponding blood prime state field ( index 6 ) causes the issue and UI is not getting updated.

- Regardless, updated the states to avoid later confusion.

    • -1
    • +5
    ./pretreatment/PreTreatmentPrimeStack.qml
DEN-15812: UI: Add an option on the Configurations setup screen to prevent the Device Registration cleanup

- Managed the Settings/System.conf owner

  1. … 3 more files in changeset.
DEN-15812: UI: Add an option on the Configurations setup screen to prevent the Device Registration cleanup

- Added the UI portion of the update

- Added the -U cmd-line option to get UI Application in update mode.

  1. … 11 more files in changeset.
DEN-15805: Updated - UI: Pressure Adjustment Screen Does Not Reset When User Cancels Change

Modify fix to use similar technique for setting and resetting in treatment flow screen

    • -23
    • +36
    ./treatment/adjustments/TreatmentAdjustmentPressuresLimits.qml
DEN-12842: Updated - UI: prepopulate option on the create treatment screen (Phase 1b)

Remove model refactoring and restored to original

Corrected slider + switch state update for prepopulating

    • -24
    • +43
    ./pretreatment/create/PreTreatmentCreate.qml
  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./endtreatment/EndTreatmentRinsebackStack.qml
    • -3
    • +3
    ./endtreatment/rinseback/EndTreatmentRinsebackReconnect.qml
    • -1
    • +1
    ./treatment/adjustments/TreatmentAdjustmentUltrafiltrationConfirm.qml
  1. … 5 more files in changeset.
DEN-15282: FW: Resume RINSBACK is missing a timeout timer

added the timer for the rinseback in progress.

    • -1
    • +0
    ./endtreatment/EndTreatmentRecirculateStack.qml
    • -66
    • +0
    ./endtreatment/rinseback/EndTreatmentRinsebaReconnect.qml
    • -0
    • +65
    ./endtreatment/rinseback/EndTreatmentRinsebackReconnect.qml
  1. … 1 more file in changeset.
DEN-15769: UI: Usability: Remove Option to increase/decrease treatment time after UF volume adjustments

    • -0
    • +1
    ./treatment/adjustments/TreatmentAdjustmentUltrafiltrationConfirm.qml
DEN-14934: Rinseback "Back to Treatment" Workflow Missing Patient Reconnection Instructions (Phase 1)

    • -0
    • +1
    ./endtreatment/EndTreatmentRecirculateStack.qml
    • -0
    • +8
    ./endtreatment/EndTreatmentRinsebackStack.qml
    • -0
    • +66
    ./endtreatment/rinseback/EndTreatmentRinsebaReconnect.qml
  1. … 5 more files in changeset.
DEN-15112: Added - UI: UI Does not Display (Pending) Treatment Logs on Export Logs Screen

WIP added the SD-Card file model to the code.

    • -23
    • +21
    ./settings/SettingsExportLogs.qml
  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./settings/SettingsDeviceRegistration.qml
    • -3
    • +3
    ./settings/SettingsManufacturingSetup.qml
    • -3
    • +3
    ./settings/SettingsServicePassword.qml
    • -2
    • +2
    ./settings/SettingsVolumeBrightness.qml
  1. … 59 more files in changeset.
Merge branch 'DEN-15279-UI-BN-S97-Cybersecurity' into staging

    • -0
    • +40
    ./settings/SettingsDecommission.qml
    • -2
    • +2
    ./settings/SettingsDeviceRegistration.qml
    • -0
    • +40
    ./settings/SettingsFactoryReset.qml
    • -0
    • +38
    ./settings/SettingsManufacturingSetup.qml
    • -0
    • +100
    ./settings/SettingsRootSSHAccess.qml
    • -0
    • +77
    ./settings/SettingsServicePassword.qml
    • -273
    • +132
    ./settings/SettingsStack.qml
    • -19
    • +6
    ./settings/SettingsVolumeBrightness.qml
  1. … 101 more files in changeset.