Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15269: Changed - UI: Vitals Keyboard Appears during Treatment Paused Alarm/State

Updated to not show keyboard when the timer times out, keyboard will show when user selects input field

UI-DEN-15530-UI-VD-S99 - Vital input bug - [ Merged ]
UI-DEN-15530-UI-VD-S99 - Vital input bug - [ Merged ]
[ On Hold - P1B ] UI-DEN-15530-UI-VD-S99 - Pre-populate TxCreate
[ On Hold - P1B ] UI-DEN-15530-UI-VD-S99 - Pre-populate TxCreate
DEN-15442: Updated - UI: Cybersecurity: Decommissioning [Settings Menu, Service Mode]

Updated the decommission qml to reflect CR feedback

change condition to show decommission and factory reset options in settings

    • -7
    • +10
    ./pages/settings/SettingsDecommission.qml
  1. … 1 more file in changeset.
DEN-15441: Updated - UI: Cybersecurity: Reset Factory [Settings Menu, Service Mode]

Update factory reset qml file based on CR feedback

    • -10
    • +3
    ./pages/settings/SettingsFactoryReset.qml
RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

RESOLVED

DEN-15361 : Changed - UI: Dialysate temp default in treatment settings is incorrect

Removed setting of value property in the pre-tx screen per CR feedback

RESOLVED

RESOLVED

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-15542 : Added : UI: Usability: Pressure Monitor window on "Additional" Saline bolus screen. Add stop cancel bolus

- Added the Arterial/Venous pressure to the Additional Rinseback.

UI-DEN-15594_UI BN S99 Usability
UI-DEN-15594_UI BN S99 Usability
DEN-15493 : Usability: UI, FW: rinseback flow rate

- removed the rinseback range slider form the create treatment screen and passing the default value of the 200 set in the configurations for the message.

- min max has also changed to 100 - 300 in ui.config repository.

  1. … 1 more file in changeset.
Thanks, Fixed.

Thanks,
Fixed.

Thanks, Fixed.

Thanks,
Fixed.

Thanks, Fixed.

Thanks,
Fixed.

Thanks, Fixed.

Thanks,
Fixed.

Thanks, Fixed.

Thanks,
Fixed.

Thanks, Fixed.

Thanks,
Fixed.

Thanks, Fixed.

Thanks,
Fixed.

  1. … 1 more file in changeset.
DEN-15400: UI: Cybersecurity: UI needs to auto generate the password for the Encrypted partition

- completed the encryption partition creation

- used a function to get the settings home, which is going to be different when App is in manufacturing mode and in normal operation.

- now in normal mode it needs to be mounted to be used and needs some movements and threading.

- For now the application should use the -R CLIS to make the app use the root home folder for the configurations, and I add it to the run.sh, but it has to be removed later.

  1. … 19 more files in changeset.
DEN-15442: Added - UI: Cybersecurity: Decommissioning [Settings Menu, Service Mode]

Added decommission.sh script

Added the invoking code of decommission script from UI

    • -0
    • +47
    ./pages/settings/SettingsDecommission.qml
  1. … 12 more files in changeset.