RESOLVED in CODE WALKTHROUGH.
Done.
Yes, they are going to be used in mode flush and chemical disinfect as well as heat disinfect.
Any reason why these are in common header? Does anybody else use them? If not, make private to HeatDisinfect module.
Done. Updated syringe pump driver to match HD-DEN-7117-1.
Good catch. Fixed.
Updated to match DEN-7091.
In comments, replace ten and five with 10 and 5.
Fixed.
It would be helpful to say which syringe pump controller signal is being passed via all these channels instead of just a channel number.
Comments start with capital letter.
Spell out lc variable as loadCell.
Align comments.