DEN-14689: updated msg_ids to be consistent with data, request, and response types.

Fixed.

Fixed.

Fixed.

Fixed.

Fixed.

Fixed.

Fixed?

Fixed?

To answer this and comment below: above_high_safety_range = temp > 46 above_low_safety_range = temp > 42 above_target_temp = target + 4 below_target_temp = target - 4 below_low_safety_range = temp...

To answer this and comment below:

above_high_safety_range = temp > 46
above_low_safety_range = temp > 42
above_target_temp = target + 4
below_target_temp = target - 4
below_low_safety_range = temp < 33

I did not see any specification regarding to separate or not to separate. Original functionality had separate triggers. Respective PRS's state OR between the two triggers.
I think it would be clear to for us in terms of debugging to see which alarm is triggered since on the low side, temp - 4 would be lower than 33, (31).

Wording may be unclear. Im up for suggestions to help clarify the alarm differences.

Fixed. Thanks!

Fixed. Thanks!

Shouldnt this be set to 0U instead of 0.0F?

Shouldnt this be set to 0U instead of 0.0F?

This should be U32.

This should be U32.

This should be U32.

This should be U32.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Should we check for stall in bolus & continuous states?

Should we check for stall in bolus & continuous states?

seems odd to create a variable just for it to be returned. Is this value supposed to be overridable?

seems odd to create a variable just for it to be returned. Is this value supposed to be overridable?

Shouldnt this be set to 0U instead of 0.0F?

Shouldnt this be set to 0U instead of 0.0F?

HD-DEN-14906_Dialyzer DE Prime
HD-DEN-14906_Dialyzer DE Prime
Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop

DEN-13154: Broke some long alarm instructions into several rows to fit on screen per DEN-14563.

Merge branch 'DEN-14830-incorrect-dialysate-temperature-alarm-triggered' into develop

Merge branch 'staging' into DEN-14763-dg_hd_dev-hd_dg_dvt-update-part-10

DEN-14830: updated alarm table to latest PRS changes regarding dialysate temperature

Merge branch 'staging' into DEN-14830-incorrect-dialysate-temperature-alarm-triggered

Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop

DEN-13154: Fixed DEN-14330 re: tilt alarms should allow blood recirculation when treatment paused.

Fixed.

Fixed.

Fixed.

Fixed.

Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm. And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm. D...

Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm.
And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm.
Did Systems want to separate these? My understanding was not to separate.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop