DENBUG64 updatd the blood detect alarms manual text

Merge branch 'DEN-14934-rinseback-back-to-treatment-workflow-missing-patient-reconnection-instructions' into develop

DEN-14934: Addressed DEN-15536.

Add resumeRinsebackRequested.

Add resumeRinsebackRequested.

Merge branch 'DEN-14934-rinseback-back-to-treatment-workflow-missing-patient-reconnection-instructions' into develop

HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
DEN-14934: Added new rinseback state to instruct user to reconnect patient before returning to treatment.

Done

Done

DEN-15642 updated the alarms table

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

See the "also" in my comment above.

See the "also" in my comment above.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Removed the extra alarm

Merged DENBUG64

Done

Done

DENBUG-64 removed the instructions of the manual alarm

Correct. Alarm 126 is not blood and dialysate recirculation.

Correct. Alarm 126 is not blood and dialysate recirculation.

Done

Done

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm? Also, since dialysate pump is on, need to start trimmer heater.

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm?

Also, since dialysate pump is on, need to start trimmer heater.

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Done

Done

Move up with other setup functions.

Move up with other setup functions.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Should be the recovery alarm.

Should be the recovery alarm.

DENBUG-64 removed alarm 63

DENBUG-64 work in progress added a new alarm for blood detection algorithm

DG-DEN-15842_Alarm 272 Fan Fault HD Alarm Does Not Declare During Post TX
DG-DEN-15842_Alarm 272 Fan Fault HD Alarm Does Not Declare During Post TX