common

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merged DENBUG-64

DENBUG-64 updated the blood detect alarm property

Merge remote-tracking branch 'origin/DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to' into develop

DENBUG-64 merged staging

DENBUG-64 updated the blood detect alarms texts

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-14934 to staging.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

It's not a fault though. User can choose to resume, rinseback or end tx but will the alarm just trigger again immediately if you haven't resolved the h/w problem? With the 2 properties set to TRUE,...

It's not a fault though. User can choose to resume, rinseback or end tx but will the alarm just trigger again immediately if you haven't resolved the h/w problem? With the 2 properties set to TRUE, user can choose to rinseback or end tx and they will not be bothered by this alarm again until rinseback and post-tx are completed.

Blaine's comment shows discrepancy between HD and DG fan handling with these flags. Would we ever want to block an HD Fan alarm? Looks to indicate a HW fail.

Blaine's comment shows discrepancy between HD and DG fan handling with these flags. Would we ever want to block an HD Fan alarm? Looks to indicate a HW fail.

Are we sure we want this?

Are we sure we want this?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

DENBUG64 updatd the blood detect alarms manual text

Merge branch 'DEN-14934-rinseback-back-to-treatment-workflow-missing-patient-reconnection-instructions' into develop

DEN-14934: Addressed DEN-15536.

Add resumeRinsebackRequested.

Add resumeRinsebackRequested.

Merge branch 'DEN-14934-rinseback-back-to-treatment-workflow-missing-patient-reconnection-instructions' into develop

HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
DEN-14934: Added new rinseback state to instruct user to reconnect patient before returning to treatment.

Done

Done

DEN-15642 updated the alarms table

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.