DG-DEN-15921_When Cap IS Open DG Transition TO Cancel Basic Path Instead OF Cancel Water Path
DG-DEN-15921_When Cap IS Open DG Transition TO Cancel Basic Path Instead OF Cancel Water Path
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

It's not a fault though. User can choose to resume, rinseback or end tx but will the alarm just trigger again immediately if you haven't resolved the h/w problem? With the 2 properties set to TRUE,...

It's not a fault though. User can choose to resume, rinseback or end tx but will the alarm just trigger again immediately if you haven't resolved the h/w problem? With the 2 properties set to TRUE, user can choose to rinseback or end tx and they will not be bothered by this alarm again until rinseback and post-tx are completed.

Blaine's comment shows discrepancy between HD and DG fan handling with these flags. Would we ever want to block an HD Fan alarm? Looks to indicate a HW fail.

Blaine's comment shows discrepancy between HD and DG fan handling with these flags. Would we ever want to block an HD Fan alarm? Looks to indicate a HW fail.

Are we sure we want this?

Are we sure we want this?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Add resumeRinsebackRequested.

Add resumeRinsebackRequested.

HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
Done

Done

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

See the "also" in my comment above.

See the "also" in my comment above.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Correct. Alarm 126 is not blood and dialysate recirculation.

Correct. Alarm 126 is not blood and dialysate recirculation.

Done

Done

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm? Also, since dialysate pump is on, need to start trimmer heater.

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm?

Also, since dialysate pump is on, need to start trimmer heater.

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Done

Done

Move up with other setup functions.

Move up with other setup functions.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Is this being permanently removed. I believe there is a requirement for generating this alarm.