DEN-15921 changed the text of alarm 126

DEN-15921 changed the alarm rank of the second blood leak alarm

Done thanks.

Done thanks.

This comment should say "bicarb pump is 2% faster than...." .

This comment should say "bicarb pump is 2% faster than...." .

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to' into staging

Merged DEN-15517

DEN-15517 added an event for testing

DG-DEN-15921_When Cap IS Open DG Transition TO Cancel Basic Path Instead OF Cancel Water Path
DG-DEN-15921_When Cap IS Open DG Transition TO Cancel Basic Path Instead OF Cancel Water Path
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DENBUG-64

DENBUG-64 updated the blood detect alarm property

Merge remote-tracking branch 'origin/DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to' into develop

DENBUG-64 merged staging

DENBUG-64 updated the blood detect alarms texts

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-14934 to staging.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

It's not a fault though. User can choose to resume, rinseback or end tx but will the alarm just trigger again immediately if you haven't resolved the h/w problem? With the 2 properties set to TRUE,...

It's not a fault though. User can choose to resume, rinseback or end tx but will the alarm just trigger again immediately if you haven't resolved the h/w problem? With the 2 properties set to TRUE, user can choose to rinseback or end tx and they will not be bothered by this alarm again until rinseback and post-tx are completed.

Blaine's comment shows discrepancy between HD and DG fan handling with these flags. Would we ever want to block an HD Fan alarm? Looks to indicate a HW fail.

Blaine's comment shows discrepancy between HD and DG fan handling with these flags. Would we ever want to block an HD Fan alarm? Looks to indicate a HW fail.

Are we sure we want this?

Are we sure we want this?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

DENBUG64 updatd the blood detect alarms manual text