This looks like pump control function. I think TD f/w should be doing this when setting Qsub before sending to DD f/w. Again, I think DD f/w should just be controlling pump to the Qsub rate given b...
This looks like pump control function. I think TD f/w should be doing this when setting Qsub before sending to DD f/w. Again, I think DD f/w should just be controlling pump to the Qsub rate given by TD f/w. HDF volume should be calculated by TD f/w by integrating commanded Qsub over time.
Generally, DD shouldn't need TMP. TD is calculating TMP. TD is using TMP to calculate Qsub. DD is just getting Qsub from TD and controlling the pump to that rate.
Generally, DD shouldn't need TMP. TD is calculating TMP. TD is using TMP to calculate Qsub. DD is just getting Qsub from TD and controlling the pump to that rate.
Doesn't match file name. Why do we have a redundant file name with ~ suffix for your new units? If these are backup files from your editor, please do not add these to the git repo (delete them befo...
Doesn't match file name. Why do we have a redundant file name with ~ suffix for your new units? If these are backup files from your editor, please do not add these to the git repo (delete them before commit).