dgfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-6402: Added flow meter check and filters for load cells

    • -9
    • +106
    /firmware/App/Controllers/LoadCell.c
    • -10
    • +3
    /firmware/App/Controllers/LoadCell.h
    • -29
    • +43
    /firmware/App/Services/Reservoirs.c
    • -25
    • +12
    /firmware/App/Services/Reservoirs.h
DG-DEN-6402_HD Dev Pretreatment Self Tests
DG-DEN-6402_HD Dev Pretreatment Self Tests
DEN-6402: Added change valve setting command

    • -0
    • +48
    /firmware/App/Services/Reservoirs.c
    • -6
    • +15
    /firmware/App/Services/Reservoirs.h
DEN-6402: Use reset persistent alarm timer

DEN-6402: Added control setup state and delay for RO pump

    • -71
    • +121
    /firmware/App/Controllers/ROPump.c
DG-DEN-5980_Non Volatile Data Management (2 of 2)
DG-DEN-5980_Non Volatile Data Management (2 of 2)
DEN-5980 disabled some functions for testing

Merge branch 'DEN-6200-hd-disposable-priming-dialyzer-side' into staging

DEN-5963 added a few lines to stop the drain pump if RPM is 0

    • -6
    • +11
    /firmware/App/Controllers/DrainPump.c
    • -3
    • +3
    /firmware/App/Controllers/Thermistors.c
DEN-5980 started adding the structs to NVDatMgmt and modified VectorCAST's configuation file to use GNU compiler template

    • -2
    • +2
    /firmware/App/Controllers/Thermistors.c
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Reverted the deletion.

Reverted the deletion.

Reverted the deletion.

Reverted the deletion.

DEN-6200: Revert heater module changes

    • -6
    • +44
    /firmware/App/Controllers/Heaters.c
Heat disinfect may need this function to set target temp to 85°C.

Heat disinfect may need this function to set target temp to 85°C.

How will DG control the trimmer heater (e.g. in heat disinfect mode) without this function? Trimmer heater not always started by HD command message.

How will DG control the trimmer heater (e.g. in heat disinfect mode) without this function? Trimmer heater not always started by HD command message.

Function has been moved to heater module. Target temperature has been added as part of the trimmer heater start command.

Function has been moved to heater module.
Target temperature has been added as part of the trimmer heater start command.

DEN-6200: Added target temperature to trimmer heater command

    • -43
    • +47
    /firmware/App/Controllers/Heaters.c
    • -3
    • +21
    /firmware/App/Controllers/Heaters.h
    • -7
    • +0
    /firmware/App/Services/MessagePayloads.h
    • -50
    • +0
    /firmware/App/Services/Reservoirs.c
Maybe new reason like dialysate temperature not set?

Maybe new reason like dialysate temperature not set?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-6200: Fixed doxygen comment

Added command id to command response.

Added command id to command response.

DEN-6200: Added command id to DG command response

    • -39
    • +59
    /firmware/App/Services/Reservoirs.c
    • -1
    • +11
    /firmware/App/Services/Reservoirs.h
The dialysate temperature HD sent to DG as another message.

The dialysate temperature HD sent to DG as another message.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.