DEN-6402: Added alarm for RO pump ramp up retry fail

In verify state, we change the target pressure with the average pressure over the verify period. That pressure value is in float. For RO pump publish data, the data type for target pressure is floa...

In verify state, we change the target pressure with the average pressure over the verify period. That pressure value is in float.
For RO pump publish data, the data type for target pressure is float as well.

For the time that the RO pump is switching between ramp and stabilize to get to the flow, the drain continues, does that affect the drain process?

For the time that the RO pump is switching between ramp and stabilize to get to the flow, the drain continues, does that affect the drain process?

Why the RPM has been lowered?

Why the RPM has been lowered?

Do we need any persistent alarm after a couple of trials? How many times are you planning to go back and forth?

Do we need any persistent alarm after a couple of trials? How many times are you planning to go back and forth?

Are we planning to provide pressure in float? Like 30.25 psi?

Are we planning to provide pressure in float? Like 30.25 psi?

Agreed. Moved to DGDefs.h

Agreed. Moved to DGDefs.h

Should this be moved to common\DG_Defs.h? I see HD has same typedef.

Should this be moved to common\DG_Defs.h? I see HD has same typedef.

DEN-6402: Added flow meter check and filters for load cells

    • -9
    • +106
    /firmware/App/Controllers/LoadCell.c
    • -10
    • +3
    /firmware/App/Controllers/LoadCell.h
    • -29
    • +43
    /firmware/App/Services/Reservoirs.c
    • -25
    • +12
    /firmware/App/Services/Reservoirs.h
DG-DEN-6402_HD Dev Pretreatment Self Tests
DG-DEN-6402_HD Dev Pretreatment Self Tests
DEN-6402: Added change valve setting command

    • -0
    • +48
    /firmware/App/Services/Reservoirs.c
    • -6
    • +15
    /firmware/App/Services/Reservoirs.h
DEN-6402: Use reset persistent alarm timer

DEN-6402: Added control setup state and delay for RO pump

    • -71
    • +121
    /firmware/App/Controllers/ROPump.c
DG-DEN-5980_Non Volatile Data Management (2 of 2)
DG-DEN-5980_Non Volatile Data Management (2 of 2)
DEN-5980 disabled some functions for testing

Merge branch 'DEN-6200-hd-disposable-priming-dialyzer-side' into staging

DEN-5963 added a few lines to stop the drain pump if RPM is 0

    • -6
    • +11
    /firmware/App/Controllers/DrainPump.c
    • -3
    • +3
    /firmware/App/Controllers/Thermistors.c
DEN-5980 started adding the structs to NVDatMgmt and modified VectorCAST's configuation file to use GNU compiler template

    • -2
    • +2
    /firmware/App/Controllers/Thermistors.c
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Reverted the deletion.

Reverted the deletion.

Reverted the deletion.

Reverted the deletion.

DEN-6200: Revert heater module changes

    • -6
    • +44
    /firmware/App/Controllers/Heaters.c
Heat disinfect may need this function to set target temp to 85°C.

Heat disinfect may need this function to set target temp to 85°C.

How will DG control the trimmer heater (e.g. in heat disinfect mode) without this function? Trimmer heater not always started by HD command message.

How will DG control the trimmer heater (e.g. in heat disinfect mode) without this function? Trimmer heater not always started by HD command message.

Function has been moved to heater module. Target temperature has been added as part of the trimmer heater start command.

Function has been moved to heater module.
Target temperature has been added as part of the trimmer heater start command.