dgfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merge branch 'DEN-14175-dg_hd_dev-hd_dg_dvt-update-part-5' into develop

DEN-14175, DEN-13999 uncommented DG pressure sensors alarm

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-14175

    • -28
    • +48
    /firmware/App/Controllers/Heaters.c
    • -3
    • +11
    /firmware/App/Services/SystemComm.c
Where is valveSensedStates being updated?

Where is valveSensedStates being updated?

Remove comment.

Remove comment.

DEN-14175, DEN-13999 updated the pressure sensors persitent alarm. Added hardware configuration functions to check the status of the device (i.e. BETA, DVT)

    • -31
    • +61
    /firmware/App/Controllers/Pressures.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Modes/ModeHeatDisinfect.c
Not clear this needs to be done. I believe is properly handled in heaters when you get a temperature value and configured for V3.

Not clear this needs to be done. I believe is properly handled in heaters when you get a temperature value and configured for V3.

DEN-14232 updated the VPo status to prevent filling the reservoirs at the end of heat disinfect

    • -8
    • +8
    /firmware/App/Modes/ModeHeatDisinfect.c
Merge branch 'staging' into DEN-14175-dg_hd_dev-hd_dg_dvt-update-part-5

DEN-14175, DIAL-213 added drain pump current and direction override commands, updated the heaters driver with the latest optimization requirements, added a DG valves sensed states override command

    • -2
    • +98
    /firmware/App/Controllers/DrainPump.c
    • -26
    • +46
    /firmware/App/Controllers/Heaters.c
    • -14
    • +95
    /firmware/App/Controllers/Valves.c
    • -30
    • +37
    /firmware/App/Controllers/Valves.h
    • -0
    • +12
    /firmware/App/Services/SystemComm.c
DEN-14232 Turn off trimmer heater before switching to FILL_R2_WITH_HOT_WATER and then back on when switching to HEAT_DISINFECT_R2_TO_R1.

    • -1
    • +9
    /firmware/App/Modes/ModeHeatDisinfect.c
DG-DEN-14232_Heat Disinfect Alarm 199 During Fill R2 With Hot Water
DG-DEN-14232_Heat Disinfect Alarm 199 During Fill R2 With Hot Water
DG-DEN-14175_DG HD Dev HD DG Dvt Update Part 5
DG-DEN-14175_DG HD Dev HD DG Dvt Update Part 5
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Modes/ModeHeatDisinfect.c
Merge branch 'DEN-13786-hd_dg_dev-s78_bb' into staging

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-13786-hd_dg_dev-s78_bb' into develop

DEN-14101: Fixed warnings on implicit declaration.

Merge branch 'staging' into DEN-13786-hd_dg_dev-s78_bb

    • -2
    • +2
    /firmware/App/Modes/ModeHeatDisinfect.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-14175-dg_hd_dev-hd_dg_dvt-update-part-5' into develop