dgfirmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

We're always ready? Why does UI need to ask if DG is ready if always ready? Replace "1" with "TRUE".

We're always ready? Why does UI need to ask if DG is ready if always ready?
Replace "1" with "TRUE".

Should this be a s/w fault?

Should this be a s/w fault?

Should default be separate from DISPENSING? And should default have a s/w fault?

Should default be separate from DISPENSING? And should default have a s/w fault?

Cycle is mis-spelled.

Cycle is mis-spelled.

The check against nearly zero is pointless since we're looking for > 10 mL/min now. Is cpTargetSpeed always positive (even when in reverse)? If not, should absolute value speed before checking agai...

The check against nearly zero is pointless since we're looking for > 10 mL/min now.
Is cpTargetSpeed always positive (even when in reverse)? If not, should absolute value speed before checking against low speed threshold.

Not really "off". Should be applied when going slowly (<10 mL/min).

Not really "off". Should be applied when going slowly (<10 mL/min).

Update function header.

Update function header.

Update function header.

Update function header.

Update function header.

Update function header.

Update function header

Update function header

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done

Done

Done

Done

Comments.

Comments.

DG-DEN-15478_Alarm 207 Occurred IN Pre Treatment Dvt 020
DG-DEN-15478_Alarm 207 Occurred IN Pre Treatment Dvt 020
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DG-DEN-13989_FW Concentrate Pump Cp1 Speed Error Observed
DG-DEN-13989_FW Concentrate Pump Cp1 Speed Error Observed
DG-DEN-15443_Alarm 97 Dialysate Conductivity Out OF Range Triggers Periodically During Treatment
DG-DEN-15443_Alarm 97 Dialysate Conductivity Out OF Range Triggers Periodically During Treatment