Merge branch 'DENBUG-318-rejection-ratio-setting-feature-implementation-in-manufacturing-configuration-for-treatment-staging' into develop

DENBUG-318 added code to keep requesting DG institutional values once start DG has been sent until the values are received

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DENBUG-318-rejection-ratio-setting-feature-implementation-in-manufacturing-configuration-for-treatment-staging' into develop

DENBUG-318 changed the min inlet water conductivity to recovery offset to 10 uS/cm. Updated the vectorcast test cases

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DENBUG-318-rejection-ratio-setting-feature-implementation-in-manufacturing-configuration-for-treatment-staging' into develop

DENBUG-318 added the request command to get the DG institutional record from HD upon starting the DG

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DENBUG-318

DENBUG-318 added an RO rejection ratio event

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DENBUG-318

DENBUG-318 changed the max RO rejection ratio to min RO rejection ratio in percent and updated the mode fill calculations to check the RR

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DENBUG-318

DENBUG-320 removed the acid and bicarb used volumes in mode fill broadcast

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DENBUG-318

DENBUG-318 removed the inlet recovery conductivity alarm #define

DENBUG-318 changed the inlet water conductivity low alarm recovery

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DENBUG-318

DENBUG-318 added a new event for min inlet water conductivity alarm from HD institutional record

DENBUG-318 updated the code to receive all the DG institutional record items. Added code to use the min inlet water conductivity alarm from the institutional record

DENBUG-319 added code to receive a command from dialing to enable the mode fill chemicals test

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DENBUG-318-rejection-ratio-setting-feature-implementation-in-manufacturing-configuration-for-treatment-staging' into develop

DENBUG-322 added code to set CPi and CPo to 20 uS/cm if the values are less than 20 uS/cm