dgfirmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DG-DENBUG-327_Baro Sensor Data Refresh Staging
DG-DENBUG-327_Baro Sensor Data Refresh Staging
Done

Done

Do we need to shift CRC bits to lsb position?

Do we need to shift CRC bits to lsb position?

DG-DENBUG-327_Baro Sensor Data Refresh Staging
DG-DENBUG-327_Baro Sensor Data Refresh Staging
DG-DEN-17299_Release 22 P1B Staging
DG-DEN-17299_Release 22 P1B Staging
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

DENBUG-318 addressed the code review comment

hd mode should be >= standby mode. You have it backward.

hd mode should be >= standby mode. You have it backward.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DENBUG-318 addressed the code review comments

Done.

Done.

DENBUG-318 updated the VectorCAST test cases

DENBUG-318 added code to request the DG institutional values in mode standby until they are received

    • -12
    • +26
    /firmware/App/Modes/ModeStandby.c
DENBUG-318 added code to keep requesting DG institutional values once start DG has been sent until the values are received

Should we have before and after values in event data?

Should we have before and after values in event data?