dgfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-14589: fixed baro sensor to consider hw config

Merge branch 'staging' into DEN-14589-DIAL-211

DEN-14604 updated the code to be able to use multiple heat disinfect timers per the new requirements

    • -96
    • +187
    /firmware/App/Modes/ModeHeatDisinfect.c
    • -1
    • +4
    /firmware/App/Modes/ModeHeatDisinfect.h
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14607 addressed the code review comments

DEN-14631 added the inlet water check functions to the POST and Standby modes

DEN-14607, DEN-13999 updated the fill mode the not command the VPo valve to fill state upon exiting the pause state. Updated the pressures driver to assume the overidden pressure is absolute and not add the baro pressure to make it absolute

    • -3
    • +10
    /firmware/App/Controllers/Pressures.c
DEN-14561 Decrease concentrate lines prime from 95 to 60 sec.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged staging with the latest updates

    • -13
    • +15
    /firmware/App/Modes/OperationModes.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Controllers/UVReactors.c
    • -1
    • +1
    /firmware/App/Controllers/UVReactors.h
    • -1
    • +1
    /firmware/App/Drivers/SafetyShutdown.c
    • -1
    • +1
    /firmware/App/Drivers/SafetyShutdown.h
    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
  1. … 8 more files in changeset.
DEN-14559 removed dead code

DEN-14559 addressed the code review comments

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
DEN-14559 addressed the code review comments

    • -13
    • +24
    /firmware/App/Modes/ModeChemicalDisinfect.c
    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
Merge branch 'staging' into DEN-14559-dg_hd_dev-hd_dg_dvt-update-part-8

    • -14
    • +17
    /firmware/App/Controllers/Pressures.c
DEN-14559 updated the code to be able request POST by the Dialin user. Updated the chemical disinfect flush to be able drain the reservoirs all the time

    • -8
    • +8
    /firmware/App/Modes/ModeHeatDisinfect.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Controllers/Thermistors.c
    • -2
    • +2
    /firmware/App/Controllers/Thermistors.h
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-13676-bicarb-low-alarm-threshold-incorrect' into staging

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-13946-certain-temperature-sensors-trigger-wrong-alarm' into staging

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-13676-bicarb-low-alarm-threshold-incorrect' into develop

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-13946-certain-temperature-sensors-trigger-wrong-alarm' into develop

    • -24
    • +24
    /firmware/App/Controllers/LoadCell.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-13946 Code review correction

    • -6
    • +1
    /firmware/App/Controllers/Thermistors.c