Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-14197: Build fixes

DEN-14197: DIAL-212: Created measured RO pump feedback voltage override. Implemented current RO Pump profile variable.

DEN-14197: Fixed warnings regarding gen idle overrides. Fixed compiler issues when merging into develop.

  1. … 1 more file in changeset.
DEN-14197: implemented RO pump profile switching. Implemented Flow to PWM function RO pump with profiles as a parameter. Created a SW fault for PI profiles out of range.

  1. … 3 more files in changeset.
DEN-14197: Initial attempt at PI controller profiles with usage in ro pump

  1. … 2 more files in changeset.
Merge branch 'DEN-14175-dg_hd_dev-hd_dg_dvt-update-part-5' into develop

  1. … 1 more file in changeset.
DEN-14175 added trimmer heater use last duty cycle to the broadcast data

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-14175 addressed the code review comments and replaced the V3 enable software configuration flag with the hardware configuration flag from FGPA. Updated the trimmer heater's ramp state to check the use previous duty cycle flag instead of estimating the duty cycle.

  1. … 5 more files in changeset.
This is done to get an average of the fill temperature as well as the last fill temperature prior to tranistioning to Gen Idle. These values are used for the next estimation and are different from ...

This is done to get an average of the fill temperature as well as the last fill temperature prior to tranistioning to Gen Idle. These values are used for the next estimation and are different from the heaters driver.

Done.

Done.

No, it should be removed.

No, it should be removed.

DEN-14175, DEN-13999 uncommented DG pressure sensors alarm

Merged DEN-14175

  1. … 3 more files in changeset.
Where is valveSensedStates being updated?

Where is valveSensedStates being updated?

Remove comment.

Remove comment.

DEN-14175, DEN-13999 updated the pressure sensors persitent alarm. Added hardware configuration functions to check the status of the device (i.e. BETA, DVT)

  1. … 2 more files in changeset.
Not clear this needs to be done. I believe is properly handled in heaters when you get a temperature value and configured for V3.

Not clear this needs to be done. I believe is properly handled in heaters when you get a temperature value and configured for V3.

Merge branch 'staging' into DEN-14175-dg_hd_dev-hd_dg_dvt-update-part-5

DEN-14175, DIAL-213 added drain pump current and direction override commands, updated the heaters driver with the latest optimization requirements, added a DG valves sensed states override command

  1. … 4 more files in changeset.
DG-DEN-14175_DG HD Dev HD DG Dvt Update Part 5
DG-DEN-14175_DG HD Dev HD DG Dvt Update Part 5
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
Merge branch 'DEN-13786-hd_dg_dev-s78_bb' into develop

Merge branch 'staging' into DEN-13786-hd_dg_dev-s78_bb

  1. … 1 more file in changeset.
Merge branch 'DEN-14175-dg_hd_dev-hd_dg_dvt-update-part-5' into develop

DEN-14175 added alarms to check the conductivity sensor's bad received hex characters