DEN-14307 updated the concentrate pumps park command

  1. … 1 more file in changeset.
Should this be removed?

Should this be removed?

Update function header.

Update function header.

Update function header.

Update function header.

Alignment

Alignment

Update function header.

Update function header.

Update function header.

Update function header.

DEN-14307 added a command to be able to start and stop heat disinfect active cool mode. Added a monitor function to check the status of the caps during the heat disinfect active cool mode. Updated heating code to use TD2 temperature sensor to set the primary heater's target for the later DVT devices

  1. … 10 more files in changeset.
DEN-14307: Added support for parking concentrate pumps.

  1. … 12 more files in changeset.
DEN-14307 updated the cleaning modes per the SRS changes

  1. … 7 more files in changeset.
Merge branch 'DEN-14433-sw-s86-mg-publish-fixes' into develop

DEN-14433: DIAL-123: fixed broadcast messages to send if we are >= of their respective intervals instead of >.

Merge branch 'DEN-13050-vpi-does-not-close-after-sample-water-returns-to-flush-idle' into develop

DEN-13757 - Remove unused function monitorLoadCellsWeightOutOfRange/

DG-DEN-13050_Vpi Does Not Close After Sample Water Returns TO Flush Idle
DG-DEN-13050_Vpi Does Not Close After Sample Water Returns TO Flush Idle
DEN-13757 - Correction to load cell alarm.

DEN-13757 - All 4 load cells have a single persistent alarm. Change to correspond to how pressure alarms are handled.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-14253 addressed the code review comments

Done.

Done.

Done.

Done.

Can't we just say isFlowLow = ( measFlow > minFlow ? FALSE : TRUE );

Can't we just say isFlowLow = ( measFlow > minFlow ? FALSE : TRUE );