DIAL-222: changed set heater duty cycle to consider overriden values on reset.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

I think VPi is closed during standby mode, so not sure why we want to check inlet conductivity in this mode. I imagine new system would get installed by service - would need some kind of priming do...

I think VPi is closed during standby mode, so not sure why we want to check inlet conductivity in this mode.
I imagine new system would get installed by service - would need some kind of priming done in service mode, so I'm not worried about that.

will these alarms trigger with open air? Ex: if a user has a device shipped to them, will they be greeted with conductivity alarms on first power up before any water has ran through the system? If...

will these alarms trigger with open air?

Ex: if a user has a device shipped to them, will they be greeted with conductivity alarms on first power up before any water has ran through the system? If so, how will they be able to fix it?

DEN-14999 updated the code based on the alarms table changes

  1. … 7 more files in changeset.
DEN-14999 updated the code based on the alarms table update

  1. … 5 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Merged DEN-14999

  1. … 5 more files in changeset.
DEN-14999 fixed the pressures self test

DEN-14999: Updated 24V range alarms.

DEN-14646 Changes per code review

Merge branch 'staging' into DEN-14646-fpga-alarm-properties-and-timing-3

DEN-14646 Update per code review

DEN-14999 merged the remote changes and updated the fill mode with the acid conductivity values

  1. … 1 more file in changeset.
I don't see it, but resolving anyway. RESOLVED in CODE WALKTHROUGH.

I don't see it, but resolving anyway.
RESOLVED in CODE WALKTHROUGH.

DG-DEN-14788_Dial 222 Sys Request
DG-DEN-14788_Dial 222 Sys Request
DIAL-222: removed heater check for reset override command.

DEN-14999: Removed alarm 203 (dg POST adc rail check for pressure sensors). Removed alarm 116 (dg temperature sensors adc rail check). Removed adc temperature sensors from alarm 115 (dg temperature sensors range check).

Spaces inside (). And explicit condition (i.e. ( TRUE == isConvNotValid ).

Spaces inside (). And explicit condition (i.e. ( TRUE == isConvNotValid ).

Corrected

Corrected

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
DEN14646 - Valves variable renamed.

Merge branch 'staging' into DEN-14646-fpga-alarm-properties-and-timing-3

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

I did. I still see a call to sendAckResponseMsg() and I don't think it should be here.

I did. I still see a call to sendAckResponseMsg() and I don't think it should be here.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.