RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

We're always ready? Why does UI need to ask if DG is ready if always ready? Replace "1" with "TRUE".

We're always ready? Why does UI need to ask if DG is ready if always ready?
Replace "1" with "TRUE".

Should this be a s/w fault?

Should this be a s/w fault?

Should default be separate from DISPENSING? And should default have a s/w fault?

Should default be separate from DISPENSING? And should default have a s/w fault?

Cycle is mis-spelled.

Cycle is mis-spelled.

The check against nearly zero is pointless since we're looking for > 10 mL/min now. Is cpTargetSpeed always positive (even when in reverse)? If not, should absolute value speed before checking agai...

The check against nearly zero is pointless since we're looking for > 10 mL/min now.
Is cpTargetSpeed always positive (even when in reverse)? If not, should absolute value speed before checking against low speed threshold.

Not really "off". Should be applied when going slowly (<10 mL/min).

Not really "off". Should be applied when going slowly (<10 mL/min).

Update function header.

Update function header.

Update function header.

Update function header.

Update function header.

Update function header.

Update function header

Update function header

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done

Done

Done

Done

Comments.

Comments.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DG-DEN-13989_FW Concentrate Pump Cp1 Speed Error Observed
DG-DEN-13989_FW Concentrate Pump Cp1 Speed Error Observed
DG-DEN-15443_Alarm 97 Dialysate Conductivity Out OF Range Triggers Periodically During Treatment
DG-DEN-15443_Alarm 97 Dialysate Conductivity Out OF Range Triggers Periodically During Treatment
Looks like it's already there.

Looks like it's already there.

Add alarmDetectedPendingTrigger to header.

Add alarmDetectedPendingTrigger to header.

DG-DEN-15455_FW Dvt 6 And 16 Cap Switches DO Not Update When Caps Close - Ready
DG-DEN-15455_FW Dvt 6 And 16 Cap Switches DO Not Update When Caps Close - Ready
DG-DEN-15477_Add Windowed Timer TO Drain Pump Encoder Error
DG-DEN-15477_Add Windowed Timer TO Drain Pump Encoder Error