-DEN-15367 fixed the transition to cancel water path in RO permeate sample mode -DEN-14004 changed the request message ID to 0x89 from UI to send the DG service record -DEN-15192 changed the concentrate pumps low speed check threshold -DEN-15509 changed the reservoir full threshold from 1700 mL to 1850 mL.

  1. … 3 more files in changeset.
DEN-14004 added another message to receive and send DG service record upon request from HD. The format of the payload is different from the format that is sent to UI

  1. … 3 more files in changeset.
Done

Done

Done

Done

DEN-15367, DEN-15389, DEN-15511, DEN-15443 added code in the mode fill to check for the conductivity of the values in the produce state and have a timeout. Removed the acid check at the end of the mode fill

  1. … 1 more file in changeset.
DEN-15367, DEN-15404 worked on RO permeate sample mode

  1. … 2 more files in changeset.
DEN-15367, DEN-15404 work in progress. Updated the RO permeate sample mode code to be able to receive and send ready command to HD for further dispensing or canceling the run

  1. … 5 more files in changeset.
DEN-15367 added test configuration to be able to run chemical disinfect without PAA acid

Comments.

Comments.

DEN-15367, DEN-15446 renamed the function that checks whether any of the inlet water conditions are active in a cleaning mode

  1. … 2 more files in changeset.
Merged DEN-15367

  1. … 6 more files in changeset.
DEN-15446 added a function to check the status of the inlet water conditions and trigger the cleaning modes to transition to cancel water path

  1. … 2 more files in changeset.
DEN-15367 merged staging

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DG-DEN-15478_Alarm 207 Occurred IN Pre Treatment Dvt 020
DG-DEN-15478_Alarm 207 Occurred IN Pre Treatment Dvt 020
Merged DEN-15478

  1. … 1 more file in changeset.
DEN-15478 updated the code to not check for the integrated volume vs. flow sensors if the fill is less than 550 mL

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-15455 updated the chemical disinfect caps monitor

DEN-15455 updated the chemical and chemical flush disinfect cap status check in start state

DEN-15367 added an RO permeate sample message

  1. … 3 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-15376 updated the chemical disinfect mode

DG-DEN-15443_Alarm 97 Dialysate Conductivity Out OF Range Triggers Periodically During Treatment
DG-DEN-15443_Alarm 97 Dialysate Conductivity Out OF Range Triggers Periodically During Treatment