We're starting seeing a trend where V&V want to override pre-calibrated values ( like the LC structure above ). I followed the LC override style since it has similar behavior. I have no issue split...

We're starting seeing a trend where V&V want to override pre-calibrated values ( like the LC structure above ). I followed the LC override style since it has similar behavior. I have no issue splitting

This is non-standard override record and approach. Is there any reason why you didn't just create a separate override command and message to override raw flow?

This is non-standard override record and approach. Is there any reason why you didn't just create a separate override command and message to override raw flow?

DG-DIAL-273_Need Override For Dg.Flow Sensors.Measured Raw Dialysate Flow Lpm
DG-DIAL-273_Need Override For Dg.Flow Sensors.Measured Raw Dialysate Flow Lpm
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List 3
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List 3
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List 2
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List 2
DG-DEN-15916_Heat Disinfect Active Cool Fails TO Start IN Start Mode
DG-DEN-15916_Heat Disinfect Active Cool Fails TO Start IN Start Mode
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done thanks.

Done thanks.

This comment should say "bicarb pump is 2% faster than...." .

This comment should say "bicarb pump is 2% faster than...." .

DG-DEN-15921_When Cap IS Open DG Transition TO Cancel Basic Path Instead OF Cancel Water Path
DG-DEN-15921_When Cap IS Open DG Transition TO Cancel Basic Path Instead OF Cancel Water Path
DG-DEN-15736_Concentrate Pump Runs IN Opposite Direction During Concentrate Self Test
DG-DEN-15736_Concentrate Pump Runs IN Opposite Direction During Concentrate Self Test