Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 86 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-201: Addressed DENBUG-201.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Fixed merge conflict.

DENBUG-201: Addressed DENBUG-201 - prevent empty concentrate alarms in post-tx mode.

DEN-16499 removed the DG usage record response from the standby mode

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-187 changed the code to set the chem disinfect start date at the prime state and changed the other DG modes to check against the chem disinfect start date

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-119 removed the transition from the service mode to ro permeate sample

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DENBUG-116 merged staging

DENBUG-116 changed the transition from service mode to fault mode to be illegal

DENBUG-124 added code to reinitialize the drain parameters chemical disinfect, chemical disinfect flush and flush in cancel water path

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-119 changed the operation modes table to prevent a mode transition from service mode to RO permeate sample mode

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-16466-alarm-282-dg-bicarbonate-concentration-bottle-low-volume-triggers-after-treatment' into develop

DEN-16446: Address DENBUG-105 to reset concentrate volume used when priming conc lines.

DENBUG-76 added code to not check the temperature sensors drift until a couple of minutes into R1 to R2

DEN-16466 Ignore bicarb/acid low volume alarms when at end of treatment or post,

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DENBUG-96: Addressed DENBUG-97.

  1. … 1 more file in changeset.
DENBUG-96: Addressed DENBUG-98.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 10 more files in changeset.
Merge branch 'DEN-16466-alarm-282-dg-bicarbonate-concentration-bottle-low-volume-triggers-after-treatment' into develop

# Conflicts:

# firmware/App/DGCommon.h

DEN-16466 Allow alarm ID 282/281 to only occur in pre-treatment or treatment modes.