DEN-9054 updated the code

  1. … 33 more files in changeset.
I believe this will eventually be a constant in the non-volatile calibration record. So will have a getter function in NV-Data module for this value.

I believe this will eventually be a constant in the non-volatile calibration record. So will have a getter function in NV-Data module for this value.

Fixed merge conflicts.

  1. … 17 more files in changeset.
Fixed.

Fixed.

Fixed.

Fixed.

Merge branch 'staging' into DEN-8103-dg-hd_dev-alarm-design

  1. … 14 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

This a matter of trying the temperature sensors more than once before we start the cycle. Making the wait time longer will not necessarily help the sensors to read within the range.

This a matter of trying the temperature sensors more than once before we start the cycle. Making the wait time longer will not necessarily help the sensors to read within the range.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Merge branch 'DEN-8103-dg-hd_dev-alarm-design' into develop

  1. … 18 more files in changeset.
CONC_PUMPS_REVERSE_SPEED_ML_PER_MIN = -30.0 to indicate that the direction is in reverse.

CONC_PUMPS_REVERSE_SPEED_ML_PER_MIN = -30.0 to indicate that the direction is in reverse.

Done.

Done.

Good catch, thanks. I changed it.

Good catch, thanks. I changed it.

I check in the inlet pressure only at the states that are needed. There are times that I am not even pulling water from the inlet line.

I check in the inlet pressure only at the states that are needed. There are times that I am not even pulling water from the inlet line.

It is no longer needed.

It is no longer needed.

2000 uS/cm is a safe value to determine if the acid is inserted in. Due to sensor to sensor variations, we decided to keep it a little lower than its conductivity value is.

2000 uS/cm is a safe value to determine if the acid is inserted in. Due to sensor to sensor variations, we decided to keep it a little lower than its conductivity value is.

Done.

Done.