Services

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DG-DEN-15016_System Faults During Treatment Then Returns TO The Standby Screen

Update for code review.

Done.

Done.

Space between ((

Space between ((

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

I think VPi is closed during standby mode, so not sure why we want to check inlet conductivity in this mode. I imagine new system would get installed by service - would need some kind of priming do...

I think VPi is closed during standby mode, so not sure why we want to check inlet conductivity in this mode.
I imagine new system would get installed by service - would need some kind of priming done in service mode, so I'm not worried about that.

HD-DEN-15016_System Faults During Treatment Then Returns TO The Standby Screen

Closing code review comments.

will these alarms trigger with open air? Ex: if a user has a device shipped to them, will they be greeted with conductivity alarms on first power up before any water has ran through the system? If...

will these alarms trigger with open air?

Ex: if a user has a device shipped to them, will they be greeted with conductivity alarms on first power up before any water has ran through the system? If so, how will they be able to fix it?

Done. Does HD communicating always account for UI traffic as well?

Done. Does HD communicating always account for UI traffic as well?

Done

Done

Reverted.

Reverted.

DEN-14999: Small fix to DG WD POST.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Merged DEN-14999

  1. … 3 more files in changeset.
I think if HD not communicating, just clear the pending ack list - when HD revives, any pending messages in here are likely moot.

I think if HD not communicating, just clear the pending ack list - when HD revives, any pending messages in here are likely moot.

I think it's better to end the if not POST body after the retries decrement (so don't use up retry during post, but still resend msg).

I think it's better to end the if not POST body after the retries decrement (so don't use up retry during post, but still resend msg).

DEN-15016-system-faults-during-treatment-then-returns-to-the-standby-screen

Updates to hold off processing retries until after INIT, and after communication with HD started.

Merge branch 'DEN-14999-alarm-table-updates' of ssh://dvm-linux02:7999/dg/dgfirmware into DEN-14999-alarm-table-updates

Fixed handling of HD cmd to change DG reservoir valve config.

DEN-14999, DEN-15030 updated the start DG command function

DEN-14999, DEN-15031 added code to receive and support different types of acid and bicarb in the treatment

  1. … 3 more files in changeset.
No need to increase this one.

No need to increase this one.

DEN-14999 merged the remote changes and updated the fill mode with the acid conductivity values

  1. … 1 more file in changeset.
I don't see it, but resolving anyway. RESOLVED in CODE WALKTHROUGH.

I don't see it, but resolving anyway.
RESOLVED in CODE WALKTHROUGH.

DEN-15016_System Faults During Treatment Then Returns TO The Standby Screen

Added a space.

DEN-15016-system-faults-during-treatment-then-returns-to-the-standby-screen

Increased MSG ACK timeouts. Delayed expecting ACKs until HD turns on and is sending messages.

DG-DEN-15016_System Faults During Treatment Then Returns TO The Standby Screen
DG-DEN-15016_System Faults During Treatment Then Returns TO The Standby Screen
Spaces inside (). And explicit condition (i.e. ( TRUE == isConvNotValid ).

Spaces inside (). And explicit condition (i.e. ( TRUE == isConvNotValid ).

Corrected

Corrected

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.