Merge branch 'DEN-15660-test_configuration_treatment_recovery' into develop

  1. … 1 more file in changeset.
DEN-15660, DEN-15767, DEN-15781 cleaned up the SystemCommMessages.c for messages list script. Added moving average for TDi and TRo sensors for dialysate temperature drift check

  1. … 2 more files in changeset.
Add nelsonSupport

Add nelsonSupport

Update header.

Update header.

Update header: rsrvr2Status, isRsrveLeadking, chemDisinfectReservoirTime, isChemDisinfectTempAboveTarget, ischemDisinfectWarmupTargetReached (should be isChemDisinfectWarmupTargetReached).

Update header: rsrvr2Status, isRsrveLeadking, chemDisinfectReservoirTime, isChemDisinfectTempAboveTarget, ischemDisinfectWarmupTargetReached (should be isChemDisinfectWarmupTargetReached).

Add stateTrialCounter

Add stateTrialCounter

Add nelsonSupport, stateTrialCounter to header.

Add nelsonSupport, stateTrialCounter to header.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DG-DEN-15701_Cannot Create Treatment After End Treatment During Pre Treatement
DG-DEN-15701_Cannot Create Treatment After End Treatment During Pre Treatement
Fixed merge conflict.

DEN-15701: handle stop DG command from HD while in standby water sampling states.

  1. … 2 more files in changeset.
DG-DEN-15660_Test Configuration Treatment Recovery
DG-DEN-15660_Test Configuration Treatment Recovery
Merged staging

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 20 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

  1. … 3 more files in changeset.
DEN-15367 removed the payload from the message that informs HD that DG is ready to dispense more fluid in DG. DEN-15571 fixed the chemical disinfect state machine to transition to cancel water path once the target temperature was not reached within the specified time

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Still has a payload. Should remove it.

Still has a payload. Should remove it.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Removed the check to make sure the target speed is > 0. The current speed is always >= 0.0 mL/min.

Removed the check to make sure the target speed is > 0. The current speed is always >= 0.0 mL/min.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH