DEN-16446 updated the code and VectorCAST tests

  1. … 1 more file in changeset.
DEN-15791: Doxygen comment fixes.

  1. … 5 more files in changeset.
DG-DEN-16446_DG Dev Test Clean UP
DG-DEN-16446_DG Dev Test Clean UP
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Dara suggested removing this, as the feature was incomplete and appears to cause odd timer/alarm behavior.

Dara suggested removing this, as the feature was incomplete and appears to cause odd timer/alarm behavior.

You are removing recover treatment test config functionality? Is this moving somewhere else or we've decided not to implement?

You are removing recover treatment test config functionality? Is this moving somewhere else or we've decided not to implement?

Merged DEN-15973

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Merge branch 'DIAL-283-no-moving-average-commands-conductivity' into develop

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Correcting issue with develop RESERVOIR_DATA_T

DIAL-283: message handling for moving average override.

  1. … 2 more files in changeset.
DG-DIAL-283_NO Moving Average Commands Conductivity
DG-DIAL-283_NO Moving Average Commands Conductivity
DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Reintroduced reservoir data structure change.

Corrected the setModeFillStateTransition semantics.

  1. … 1 more file in changeset.
DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Structure out of sync between HD/DG. Corrected.

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

Modification to test DG comm issues.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Add blank line between this code and switch below.

Add blank line between this code and switch below.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

These fields have been moved to a different broadcast at some point, and were no longer being set by the reservoir data broadcast. The fields regularly contained "NAN" values, so I removed them.

These fields have been moved to a different broadcast at some point, and were no longer being set by the reservoir data broadcast. The fields regularly contained "NAN" values, so I removed them.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Why are these fields removed?

Why are these fields removed?

Done. Code updates pushed.

Done. Code updates pushed.

DEN-16314-1-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low

fill time allowance for time spent in pause state

Cleared fill timer on transition to generate

Added removed unused (and uninitialized) variables from broadcast data structure.