RESOLVED in CODE WALKTHROUGH.
Need to have 2 blank lines.
Need to add a check if switchId is in valid range.
Do we need to add copyrights or bamboo will do it?
Common has an enumeration (OPN_CLS_STATE_T) that is similar to this. Suggest using from common.
Should this function move to ModeRecirculate? It does not seem to contain any information related to drain pump.
What happens when calibration data is invalid?
The comment is out-date and need to be updated.
Param name is not matched.
This can fit in one line.
Should it be 0 and is the first one in the enumeration?
Done.
This is fixed in the latest commit.
The default pump control mode is none.