DEN-15367, DEN-15446 renamed the function that checks whether any of the inlet water conditions are active in a cleaning mode

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-15367

DEN-15446 added a function to check the status of the inlet water conditions and trigger the cleaning modes to transition to cancel water path

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-15367 merged staging

    • -3
    • +12
    ./Modes/ModeHeatDisinfectActiveCool.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DG-DEN-15478_Alarm 207 Occurred IN Pre Treatment Dvt 020
DG-DEN-15478_Alarm 207 Occurred IN Pre Treatment Dvt 020
Merged DEN-15478

DEN-15478 updated the code to not check for the integrated volume vs. flow sensors if the fill is less than 550 mL

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-15455 updated the chemical disinfect caps monitor

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-15455 updated the chemical and chemical flush disinfect cap status check in start state

DEN-15367 added an RO permeate sample message