DEN-15571, DEN-15398 changed the fill mode full speed prime time from 8 seconds to 7 seconds

DEN-15571 addressed the code review comments

Done

Done

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-15571 updated the chemical disinfect drain status function

Several variables missing from header.

Several variables missing from header.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-15571-chemical-disinfect-rx-to-ry-timeout-alarms-do-not-trigger' into develop

DEN-15443 added code to transition back to ramp state in concenrate pumps if the new requested target speed is greater than the threshold

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-15571 made changes in the chemical disinfect mode to not trigger the max temeprature and conductivity out of range until the end of the chemical disinfect mode

DG-DEN-15571_Chemical Disinfect RX TO RY Timeout Alarms DO Not Trigger
DG-DEN-15571_Chemical Disinfect RX TO RY Timeout Alarms DO Not Trigger
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged staging

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
Merge branch 'staging' into DEN-15367-hd_dg_test_configuration_setup

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-14657 removed code to not check the error bit of a temperature sensor and let the temperature out of range alarm trigger the alarm

Done

Done

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-15571 changed the chemical disinfect temperature timeout from 1 minute to 5 minutes

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

DEN-15367 removed the payload from the message that informs HD that DG is ready to dispense more fluid in DG. DEN-15571 fixed the chemical disinfect state machine to transition to cancel water path once the target temperature was not reached within the specified time

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.