firmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DG-DEN-15027_Heat Disinfect Transfer Hot Water Lower Flow
DG-DEN-15027_Heat Disinfect Transfer Hot Water Lower Flow
Can we rename variable to reflect it is a timer counter at least?

Can we rename variable to reflect it is a timer counter at least?

The variable that define is being compared to is valveStateMismatchCounter which is why I called it such.

The variable that define is being compared to is valveStateMismatchCounter which is why I called it such.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Yes, and #define has "COUNT" in its name which was my point.

Yes, and #define has "COUNT" in its name which was my point.

DG-DEN-15014_DG HD Dev HD DG Dvt Update Part 11
DG-DEN-15014_DG HD Dev HD DG Dvt Update Part 11
Corrected

Corrected

Alarm is based on a counter. The #define corresponds to a 100ms time.

Alarm is based on a counter. The #define corresponds to a 100ms time.

Corrected

Corrected

Added DEBUG WARNING

Added DEBUG WARNING

DG-DEN-14999_Alarm Table Updates
DG-DEN-14999_Alarm Table Updates
DG-DEN-14788_DIAL-222
DG-DEN-14788_DIAL-222
DG-DEN-14817_Concentrate Pump Triggers Alarm 187 188 After Pump Parking
DG-DEN-14817_Concentrate Pump Triggers Alarm 187 188 After Pump Parking
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Done

Done