firmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./App/Controllers/ConductivitySensors.c
DEN-15375, DEN-15374, DEN-15378 fixed release build software fault

    • -1
    • +2
    ./App/Controllers/ConductivitySensors.c
DEN-15375 updated the heaters control

DEN-15375 updated the code

DEN-15375 updated the code

DEN-15375 addressed the heaters fixes

DG-DEN-15375_Dialysate Heating Fixes
DG-DEN-15375_Dialysate Heating Fixes
DEN-15367 worked on the treatment recovery

    • -0
    • +24
    ./App/Services/SystemCommMessages.c
DG-DEN-15369_FW DG Framing Error Generates Incorrect Alarm
DG-DEN-15369_FW DG Framing Error Generates Incorrect Alarm
DEN-15369 Correct framing error alarm ID

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged staging

    • -1
    • +1
    ./App/Controllers/ConductivitySensors.c
DG-DEN-15367_HD DG Test Configuration Setup
DG-DEN-15367_HD DG Test Configuration Setup
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./App/Controllers/ConductivitySensors.c
    • -1
    • +1
    ./App/Controllers/ConductivitySensors.h
    • -1
    • +1
    ./App/Controllers/TemperatureSensors.c
    • -1
    • +1
    ./App/Modes/ModeChemicalDisinfectFlush.c
    • -1
    • +1
    ./App/Modes/ModeHeatDisinfectActiveCool.c
  1. … 5 more files in changeset.
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'DEN-15274-hd_dg_test_configuration_setup' into develop

DEN-15274 changed the target flow of the fill pause to 0.8 L/min

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-15274

Done.

Done.

DEN-15274 addressed the code review comments

Purpose is to account for cooling of dialysate from reservoir to dialyzer. The faster Qd is, the less cooling will occur (less time to cool). RESOLVED in CODE WALKTHROUGH.

Purpose is to account for cooling of dialysate from reservoir to dialyzer. The faster Qd is, the less cooling will occur (less time to cool).
RESOLVED in CODE WALKTHROUGH.