DEN-15921, DEN-15868 changed the direction of the pumps in RO permeate sample mode. Changed the code to check the flow in heat disinfect and active cool to only when the system is not hot

    • -3
    • +2
    ./App/Modes/ModeHeatDisinfectActiveCool.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-15517

    • -4
    • +48
    ./App/Controllers/ConcentratePumps.c
DEN-15517 added an event to monitor the inactive data

DEN-15921 changed the active cool to transition to cancel water path instead of basic path when caps are not on

    • -1
    • +1
    ./App/Modes/ModeHeatDisinfectActiveCool.c
DG-DEN-15714_EE Alarm 84 HD Rtc Time After Cycling DG And HD Dvt 02
DG-DEN-15714_EE Alarm 84 HD Rtc Time After Cycling DG And HD Dvt 02
DEN-15714 Temporarily remove synching of DG RTC from HD RT.

DG-DEN-15921_When Cap IS Open DG Transition TO Cancel Basic Path Instead OF Cancel Water Path
DG-DEN-15921_When Cap IS Open DG Transition TO Cancel Basic Path Instead OF Cancel Water Path
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    ./App/Controllers/TemperatureSensors.c
DEN-1636 removed a blank line

DEN-15736 updated the concentrate pumps to set the control bit if it is not the same as what it should be

    • -3
    • +48
    ./App/Controllers/ConcentratePumps.c
DEN-15416 Process baro sensor when not in DG_MODE_INIT

    • -3
    • +1
    ./App/Controllers/TemperatureSensors.c
DG-DEN-15416_DG Fault Baro Cal Crc ID 233 Declares After Cycling Power Dvt 020
DG-DEN-15416_DG Fault Baro Cal Crc ID 233 Declares After Cycling Power Dvt 020
DG-DEN-15736_Concentrate Pump Runs IN Opposite Direction During Concentrate Self Test
DG-DEN-15736_Concentrate Pump Runs IN Opposite Direction During Concentrate Self Test
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-15714 Correct baro sensor CRC check. Remove BETA reference. Check now runs in both RELEASE and DEBUG builds.

    • -3
    • +1
    ./App/Controllers/TemperatureSensors.c
DG-DEN-15867_System Does Concentrate Self Test Twice During Pre Treatment
DG-DEN-15867_System Does Concentrate Self Test Twice During Pre Treatment
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

DEN-15867: Fix for issue where concentrate prime/test was being done twice in pre-tx.